Skip to content

Model note · Centrifuge

Eppendorf Centrifuge 5702

The 5702 is a compact benchtop low speed centrifuge used in clinical and research labs for blood tubes, cell work and routine separations, with swing bucket and fixed angle rotors. The 5702 R and RH add refrigeration. It has an automatic imbalance detector, motorised lid lock with emergency release, two program keys and a parameter lock, and shows short display codes for faults.

The model

Maker, model and type.

Maker
Eppendorf SE
Model
Centrifuge 5702 (and 5702 R / 5702 RH)
Device type
Centrifuge
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Maximum speed and RCF4,400 rpm; 3,000 x gSource
Maximum load and kinetic energy4 x 90 mL; 2,280 Nm; sample density up to 1.2 g/mLSource
Power230, 120 or 100 V, 50 to 60 Hz; 200 W (5702), 380 W (5702 R/RH); fuses 2.5 A time lag at 230 V, 5.0 A at 120/100 VSource
Acceleration and brakingStart up and run down each under about 25 sSource
Size and weight (5702)241 x 320 x 395 mm (H x W x D); 20 kg without rotor; refrigerated versions 36 kgSource
Environment and noiseAmbient 2 to 40 C (5702), 10 to 40 C (R/RH); up to 75% RH; up to 2000 m; under 58 dB(A)Source
Fixed angle rotor bore limits3.75 g per bore on F-45-24-11; 8.7 g on F-45-18-17-CryoSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Load rotors and buckets symmetrically with only the specified tubes; imbalance increases drive wear and noise. Source
  • Wait for the rotor to stop before opening the lid; the lid stays locked while the rotor turns. Source
  • An overspeed error (Er 7) means the speed system cannot be trusted; do not keep using the unit. Source

Sources

Where this note comes from.

Not covered: The manual copy is hosted by NIST and its revision date was not checked against Eppendorf's current edition. No FDA 510(k), recall or MAUDE records were found for Eppendorf. Er 24's meaning depends on when it appears and is summarised only under failing parts.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.