Model note · Intra-aortic Balloon Pump

Arrow AutoCAT 2 series intra-aortic balloon pump, including AutoCAT 2 WAVE

The Arrow AutoCAT 2 is the IABP generation before Teleflex's AC3, and many units are still in hospital fleets. It merged four earlier Arrow pump technologies into one series with automatic timing and triggering (AutoPilot) and a manual Operator mode. The WAVE versions add fibre optic pressure sensing. It shares its therapy algorithms, bellows drive and battery recall history with the AC3, so the two are often serviced side by side.

The model

Maker, model and type.

Maker
Arrow International, Inc. (Teleflex Incorporated)
Model
AutoCAT 2 series intra-aortic balloon pump, including AutoCAT 2 WAVE (catalogue IAP-0400, IAP-0435, IAP-0500, IAP-0535)
Device type
Intra-aortic Balloon Pump
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA clearanceK060309, AutoCAT Intra-Aortic Balloon Pump Series, cleared 6 April 2006; 21 CFR 870.3535, product code DSP, now Class IISource
Predecessor technologiesAutoCAT (K983866), ACAT 1 (K965209), ACAT 2 (K002256) and the fibre optic sensor IAB black box (K021462)Source
Pneumatic driveStepper motor driven bellows with augmented deflation and no disposable partsSource
Pumping volume and balloons0.5 to 50 cc in 0.5 cc steps; for 30, 35, 40 and 50 cc balloonsSource
BatterySealed lead acid, at least 90 minutes, with an optional upgrade to double battery lifeSource
Power85 to 265 V, 47 to 63 Hz; 225 W or 3.1 ASource
Display and controls10.3 inch colour LCD with 26 backlit hard keys and no touchscreenSource
EnvironmentOperating 5 to 45 degrees C (5 to 35 degrees C with fibre optic sensor); storage -15 to 40 degrees C, 200 to 1060 hPa, 15 to 80% humiditySource
Software level in 2006 clearanceV2.22, with changes to AutoPilot timing, triggering, AP zero and calibration, purge cycle and helium refill, and alarm handlingSource

Recalls

Known FDA recalls.

  • Z-0426-20232022-10-17

    Potential short battery run times (AutoCAT 2 IAP-0400; related records cover other regional variants).

    FDA record
  • Z-2408-20202020-05-20

    Motor connector wires susceptible to vibration damage, with System Error 3 and High Baseline alarms.

    FDA record
  • Z-0550-20082007-11-12

    Possible leak in the helium drive system (IAP-0400, IAP-0435, IAP-0500, IAP-0535).

    FDA record

Safety

Safety warnings from the record.

  • A motor wiring fault can stop counterpulsation suddenly or prevent start up. Source
  • The pump is intended for transport within and between hospitals, so battery capacity is a direct patient safety item. Source

Sources

Where this note comes from.

Not covered: Teleflex no longer has a dedicated AutoCAT 2 page, so the manufacturer link points to its successor. Most specifications come from the AutoCAT 2 column of the AC3 510(k) comparison table. Which catalogue numbers are WAVE versions was not confirmed. The operator and service manuals are not public, so only alarms named in recalls are listed. Teleflex may have announced end of service dates for AutoCAT 2; none were found in an opened source. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.