Model note · Intra-aortic Balloon Pump

Arrow AC3 Optimus intra-aortic balloon pump

The AC3 Optimus is Teleflex's current intra-aortic balloon pump. It is a touchscreen successor to the Arrow AutoCAT 2 and keeps the same core therapy algorithms. A stepper motor driven bellows shuttles helium into the balloon in diastole and pulls it out before systole. The Optimus version adds the FiberOptix sensor and the WAVE IntraBeat algorithm, which times inflation beat by beat, and an AutoPilot mode chooses trigger and timing automatically.

The model

Maker, model and type.

Maker
Arrow International, LLC (Teleflex Incorporated)
Model
AC3 Optimus intra-aortic balloon pump (IAP-0700 / IAP-0701; non fibre optic AC3 IAP-0600 / IAP-0601)
Device type
Intra-aortic Balloon Pump
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
PneumaticsStepper motor driven bellows; USP grade helium; disposable canister IAH-09045 (500 psi) or refillable cylinder IAH-09048 (2000 psi)Source
Pumping volume and rate0.5 to 50 cc in 0.5 cc steps; 40 to 200 pulsations per minute; assist ratios 1:1, 1:2, 1:4, 1:8Source
Compatible balloons30, 35, 40 and 50 cc Arrow IAB catheters (UltraFlex, Ultra 8, NarrowFlex, RediGuard)Source
Electrical90 to 264 VAC, 47 to 63 Hz; typical 3.7 A at 110 VAC, maximum 5.1 A at 110 VAC and 2.8 A at 240 VACSource
BatteryAt least 90 minutes on full charge, 180 minutes with optional second battery; about 4 hours to recharge to 80%Source
Size and weightControl module 26 x 36.8 x 5.0 cm, 2.7 kg; pneumatic unit 84.6 x 33 x 62.2 cm, 44.5 kg; total 47.2 kgSource
Display13.3 inch 1208 x 800 resistive touchscreen with 3 waveform channels (ECG, arterial pressure, balloon pressure)Source
Trigger modesECG (Pattern, Peak, AFib; 50 microvolt minimum), arterial pressure (3 mmHg minimum pulse pressure), pacer (VPace, APace)Source
Operating temperature0 to 45 degrees C without fibre optic sensor, 0 to 35 degrees C with itSource
FDA classificationIntra-aortic balloon and control system, 21 CFR 870.3535, Class II, product code DSP; K162820 (decided 31 March 2017, predicate AutoCAT 2 K060309)Source

Recalls

Known FDA recalls.

  • Z-0438-20232022-10-17

    Potential short battery run times (AC3 Optimus IAP-0700; related records cover IAP-0701, AC3 and AutoCAT 2).

    FDA record
  • Z-2416-20202020-05-20

    Motor connector wires vulnerable to vibration damage, with System Error 3 and High Baseline alarms and possible sudden stop.

    FDA record
  • Z-2417-20202020-05-20

    Same motor connector issue for AC3 Optimus IAP-0701.

    FDA record

Safety

Safety warnings from the record.

  • A motor wiring fault can cause abrupt loss of counterpulsation or stop the pump starting. Source
  • Short battery life puts patients at risk during transport between departments or facilities, which is part of the intended use. Source
  • The fibre optic sensor lowers the maximum operating temperature to 35 degrees C. Source

Sources

Where this note comes from.

Not covered: The operator manual (IAM-9010) and service manual are not publicly posted, so only alarms named in FDA recalls are listed and their full meanings are not confirmed. Later 510(k)s (K192238, K201112, K232343) were not reviewed in detail. The EMEA product page was not opened directly; specifications come from Teleflex's spec sheet and the FDA summary. MAUDE counts are reports, not failure rates. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.