Siemens Healthineers Ysio Max Digital Radiography System

Ysio Max is a ceiling mounted stationary digital radiography room for whole body radiography and tomography of children, adults and bariatric patients, including emergency work. A Polydoros generator drives an OPTITOP tube on a ceiling rail with a motorised collimator, and fixed and wireless flat panel detectors feed the syngo FLC imaging system. It is not for mammography.

The model

Maker, model and type.

Maker
Siemens Healthcare GmbH (US: Siemens Medical Solutions USA, Inc.)
Model
Ysio Max Digital Radiography System
Device type
X-ray
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Generator and tubePolydoros 65/80 kW generator; OPTITOP 150/40/80/HC-100 tubeSource
Collimation and doseDigital multileaf collimator; Kerma X air kerma measurementSource
DetectorsTrixell Pixium MAX wi-D (3543EZh) and MAX mini (2430EZ) wireless; MAX static (4343RC/RCE) fixedSource
Fixed detector (4343RCE)2860 x 2874 pixels, 148 um pitch; a-Si with CsI; 16 bit; DQE 67% at 0.05 lp/mm; MTF 62% at 1 lp/mmSource
Software (VF10)Windows 10 based, added cybersecurity, extra paediatric programmes, virtual machine for hospital IT, Diamond View Plus processing for gridless exposuresSource
Model number10762470Source
FDA clearanceK181279 (June 2018, VF10), predicate K133259; product code KPR, Class II, 21 CFR 892.1680Source

Recalls

Known FDA recalls.

  • Z-2758-20202020-01-08

    Software issues including incorrectly assigned images on Ysio Max and override collision risk on related systems.

    FDA record
  • Z-1339-20202020-01-08

    Software bug could assign one image to two patients.

    FDA record
  • Z-0300-20192018-08-06

    3D V tube stand lifting column cables could break without triggering the safety lock.

    FDA record
  • Z-0015-20172016-08-01

    Patient table internal part could detach and tilt the table top; use stopped pending inspection.

    FDA record
  • Z-1667-20152015-04-13

    VE10E: studies and images could be deleted; sporadic portable detector image recovery problems.

    FDA record
  • Z-0994-20152014-12-15

    VE10 with SmartOrtho: composed images could be flipped before sending to PACS.

    FDA record
  • Z-0002-20152014-09-02

    RIS worklist related malfunction could lose images on certain serial numbers.

    FDA record

Safety

Safety warnings from the record.

  • A broken lifting column cable can let the tube arm drop unexpectedly during vertical movement. Source
  • Flipped composed images may show R/L labels wrongly and lead to misdiagnosis. Source
  • A software bug could assign one image to two patients. Source
  • Not for mammography. Source

Sources

Where this note comes from.

Not covered: Siemens operator and service documentation is not public, so there are no error codes, kV/mAs ranges, heat capacity, room dimensions or QA procedures. Recall records were read through openFDA.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.