GE HealthCare Definium 8000 digital radiographic system

The Definium 8000 is a fixed room digital radiography system launched by GE at RSNA 2005, successor to the Revolution XR/d. It pairs a ceiling mounted, motorised autopositioning tube support with amorphous silicon flat panel detectors in a wall stand and, in the full room version, under a floating tabletop. Options include Auto Image Paste for long spine and leg views and VolumeRAD digital tomosynthesis, which makes it a general purpose room for busy imaging departments.

The model

Maker, model and type.

Maker
GE HealthCare (GE Medical Systems), Waukesha, Wisconsin
Model
Definium 8000 digital radiographic system
Device type
X-ray
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Detector technologyAmorphous silicon flat panel, shared with the Definium AMX 700 portableSource
Tube suspensionOverhead tube on ceiling rails with touchscreen, motorised five axis autopositioning and a screen rotating 90 degreesSource
ConfigurationsWall stand only (one detector) or full room with fixed table and wall stand (two detectors); all versions need ceiling railsSource
Auto Image PasteJoins up to five automatically positioned exposures into one image of spine or legsSource
VolumeRADLow dose projection sweep reconstructed into multiple tomographic slices (chest, abdomen, extremities, spine)Source
Model numbers seen in FDA records5131070 and 5135678Source
Regulatory basisCleared as the Revolution XR/d family, e.g. K051967 (with tomosynthesis), 21 CFR 892.1680 stationary x-ray system, Class II, product codes KPR, MQB, IZFSource
Pre-installation manualGE document 5137435-100 listed on GE site planningSource

Recalls

Known FDA recalls.

  • Z-0584-20122011-11-22

    Patient barrier rotating arms on Definium 8000 and Discovery XR650 image pasting systems may fail to hold position.

    FDA record
  • Z-0528-20102009-08-13

    Tabletops could move unexpectedly in both directions when brakes failed to engage after pedal release.

    FDA record
  • Z-1822-20082008-03-07

    VolumeRAD processing software offset wall stand slices by 17 mm.

    FDA record
  • Z-1882-20082008-03-07

    Generator software could report higher mAs than actually delivered.

    FDA record
  • Z-1883-20082008-01-02

    Manual switch for disabling positive beam limitation on the tube support was not labelled as required.

    FDA record
  • Z-1476-20082007-08-13

    Required certification labels missing from the x-ray control.

    FDA record
  • Z-1789-20122006-05-31

    Operator's finger pinched between the fixed roller and gliding tabletop of the table.

    FDA record

Safety

Safety warnings from the record.

  • A tabletop that floats without the pedal pressed can move a patient unexpectedly during transfer and lead to a fall. Source
  • Patient barrier arms used to steady patients during image pasting may swing freely if friction is lost, affecting patient and user safety. Source
  • VolumeRAD wall stand slices were offset by 17 mm before correction, so anatomy within 17 mm of the detector could be missed. Source
  • A generator software error could give light images while reporting more mAs than was delivered, which affects dose records. Source

Sources

Where this note comes from.

Not covered: GE's operator and service manuals, generator and tube ratings, detector size and error codes for the Definium 8000 are not publicly available, so specs are limited to press and FDA descriptions and errorCodes is empty. The system is a legacy product with no current GE product page; GE's public site planning drawings were checked but carry installation layouts rather than performance specifications. Recalls for the Definium 5000, AMX 700 and 656HD were excluded as different products.