Philips DigitalDiagnost C90 Digital Radiography System

DigitalDiagnost C90 is Philips' high end ceiling suspended digital radiography room. A tube with a control handle rides on a ceiling suspension over a height adjustable table and vertical stands, with a fixed Pixium detector and wireless SkyPlate detectors run from the Eleva Workspot. It covers routine radiography including ICU, trauma and paediatrics, but not fluoroscopy, angiography or mammography.

The model

Maker, model and type.

Maker
Philips Medical Systems DMC GmbH
Model
DigitalDiagnost C90 Digital Radiography System
Device type
X-ray
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
DetectorsFixed RAD detector Pixium 4343RCE; wireless SkyPlate family, plus SkyPlate E from K202564Source
Height adjustable table (TH-2)51.5 to 91.5 cm motorised; table 335 kg; patient up to 375 kg static centre, 318 kg dynamic centre, 210 kg dynamic off centreSource
TH-2 table top240 x 75 cm floating sandwich top; travel plus or minus 56 cm longitudinal and 12.8 cm transverse; electromagnetic brakesSource
Single side suspended table (TH-S)51 to 91 cm; 214 kg; patient 225 kg at centre, 135 kg at end; 260 x 75 cm Kevlar overlay topSource
WorkstationEleva Workspot with SkyFlow image chain and workflowSource
Model numbers712034 (High Performance); 712035 (Flex/Value/Chest/ER)Source
FDA clearanceK202564 (September 2020, SkyPlate E added), predicate K182973; product codes MQB, KPR, LLZ; 21 CFR 892.1680, Class IISource

Recalls

Known FDA recalls.

  • Z-2354-20262026-05-01

    Flex/Value/Chest/ER systems delivered without the seismic kit for wall stand VS2.

    FDA record
  • Z-0465-20262025-10-01

    Ceiling suspension telescopic carriages made July 2024 to June 2025 may be hard to move vertically.

    FDA record
  • Z-0275-20242023-09-21

    Ceiling suspended units could fail and fall (also Z-0276 to Z-0278-2024 for other configurations).

    FDA record
  • Z-2042-20232023-05-19

    Software login issue may prevent users logging in.

    FDA record
  • Z-0239-20222021-09-22

    Eleva monitor missing required radiation warning and certification labels.

    FDA record
  • Z-1328-20222021-05-19

    Wall stand VS2 firmware could rotate the first image after restart (also Z-1329-2022).

    FDA record

Safety

Safety warnings from the record.

  • Ceiling suspended units could fail and fall onto staff or patients; Philips' notice must be kept with the IFU. Source
  • Wrong image orientation on the first exam could mislead diagnosis. Source
  • Not for fluoroscopy, angiography or mammography. Source
  • If carriage movement needs more than 110 N, the motor drive can report errors and restrict motorised movement. Source

Sources

Where this note comes from.

Not covered: Philips IFU and service documents are not public, so generator and tube ratings, detector performance and error codes (other than the drive error described in a recall) are missing. MAUDE counts cover the whole DigitalDiagnost brand, including the C50 and older rooms. Recall records were read through openFDA.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.