Siemens Healthineers RAPIDPoint 500e Blood Gas System

The RAPIDPoint 500e is a cartridge based point of care and laboratory blood gas analyser that measures blood gases, electrolytes, glucose, lactate and CO-oximetry including neonatal bilirubin. It uses a measurement cartridge holding the sensor module and sample probe, a separate wash/waste cartridge and optional automatic QC. The 500e is an updated RAPIDPoint 500: Windows 10 IoT replaced Windows 7 Embedded, main board parts were changed for obsolescence, 2D bar code scanning was added and QC controls were tightened.

The model

Maker, model and type.

Maker
Siemens Healthcare Diagnostics Inc.
Model
RAPIDPoint 500e Blood Gas System
Device type
Blood Gas Analyser
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ParameterspH, pCO2, pO2, Na, K, Ca, Cl, glucose, lactate, tHb, sO2, O2Hb, HHb, COHb, MetHb, neonatal total bilirubinSource
Sample volume100 uLSource
Time to resultsAbout 60 secondsSource
Measurement cartridge28 days on board or maximum test count; about 24 minutes to initialise; includes new sensor module and sample probeSource
Quality controlAutomatic or ampoule QC, three levels of independent QC solutions, customisable scheduleSource
Operating system and connectivityWindows 10 IoT; LIS, wireless, Ethernet and serial ports; POCcelerator data managementSource
Lactate method and rangeAmperometric lactate oxidase sensor; reportable 0.18 to 30 mmol/L; 2 point calibration from on board reagentSource
FDA basisK192240 Special 510(k) modifying the RAPIDPoint 500 (K122539); 21 CFR 862.1120, product code CHL, Class IISource

Recalls

Known FDA recalls.

  • Z-2230-20252025-06-26

    Wash/waste cartridge lot WW/08925 manufacturing error causing positive and negative bias across several analytes.

    FDA record
  • Z-2214-20242024-05-17

    Software v5.3 labelled capillary samples as arterial at the LIS.

    FDA record
  • Z-0606-20242023-10-26

    Lactate measurement cartridges could cause negative sodium bias and Question Result flags on electrolytes (also Z-0607 to Z-0609-2024).

    FDA record
  • Z-0422-20242023-09-26

    Drug interference from perhexiline maleate or atomoxetine causing falsely high sodium.

    FDA record

Safety

Safety warnings from the record.

  • Perhexiline maleate or atomoxetine can falsely raise sodium on RAPIDPoint 500 and 500e, risking missed hyponatraemia or wrong treatment. Source
  • Certain lactate cartridges could bias sodium low, which near severe thresholds may lead to harmful sodium correction. Source
  • In capillary mode, software v5.3 sent results to the LIS labelled as arterial, so they could be judged against arterial ranges. Source
  • Check the right patient ID is scanned before confirming the sample, as positive identification is a core training point. Source

Sources

Where this note comes from.

Not covered: A copy of the RAPIDPoint 500e Operator's Guide is hosted by a Singapore hospital group but sat behind a bot check and could not be opened, so its error code tables, maintenance schedule, dimensions and power data are not included; only messages named in FDA records are listed. Cleaning and sample handling checks come from one hospital's training sheet. There are 18 recall records for the RAPIDPoint 500 family; older 500 only items (patient ID and data stream issues, 2012 to 2018) are left out. MAUDE counts mix the 500 and 500e and are trends only.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.