Model note · Blood Gas Analyser
Siemens Healthineers RAPIDPoint 500e Blood Gas System
The RAPIDPoint 500e is a cartridge based point of care and laboratory blood gas analyser that measures blood gases, electrolytes, glucose, lactate and CO-oximetry including neonatal bilirubin. It uses a measurement cartridge holding the sensor module and sample probe, a separate wash/waste cartridge and optional automatic QC. The 500e is an updated RAPIDPoint 500: Windows 10 IoT replaced Windows 7 Embedded, main board parts were changed for obsolescence, 2D bar code scanning was added and QC controls were tightened.
The model
Maker, model and type.
- Maker
- Siemens Healthcare Diagnostics Inc.
- Model
- RAPIDPoint 500e Blood Gas System
- Device type
- Blood Gas Analyser
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Parameters | pH, pCO2, pO2, Na, K, Ca, Cl, glucose, lactate, tHb, sO2, O2Hb, HHb, COHb, MetHb, neonatal total bilirubin | Source |
| Sample volume | 100 uL | Source |
| Time to results | About 60 seconds | Source |
| Measurement cartridge | 28 days on board or maximum test count; about 24 minutes to initialise; includes new sensor module and sample probe | Source |
| Quality control | Automatic or ampoule QC, three levels of independent QC solutions, customisable schedule | Source |
| Operating system and connectivity | Windows 10 IoT; LIS, wireless, Ethernet and serial ports; POCcelerator data management | Source |
| Lactate method and range | Amperometric lactate oxidase sensor; reportable 0.18 to 30 mmol/L; 2 point calibration from on board reagent | Source |
| FDA basis | K192240 Special 510(k) modifying the RAPIDPoint 500 (K122539); 21 CFR 862.1120, product code CHL, Class II | Source |
Recalls
Known FDA recalls.
- Z-2230-20252025-06-26
Wash/waste cartridge lot WW/08925 manufacturing error causing positive and negative bias across several analytes.
FDA record - Z-2214-20242024-05-17
Software v5.3 labelled capillary samples as arterial at the LIS.
FDA record - Z-0606-20242023-10-26
Lactate measurement cartridges could cause negative sodium bias and Question Result flags on electrolytes (also Z-0607 to Z-0609-2024).
FDA record - Z-0422-20242023-09-26
Drug interference from perhexiline maleate or atomoxetine causing falsely high sodium.
FDA record
Safety
Safety warnings from the record.
- Perhexiline maleate or atomoxetine can falsely raise sodium on RAPIDPoint 500 and 500e, risking missed hyponatraemia or wrong treatment. Source
- Certain lactate cartridges could bias sodium low, which near severe thresholds may lead to harmful sodium correction. Source
- In capillary mode, software v5.3 sent results to the LIS labelled as arterial, so they could be judged against arterial ranges. Source
- Check the right patient ID is scanned before confirming the sample, as positive identification is a core training point. Source
Sources
Where this note comes from.
- Siemens Healthineers RAPIDPoint 500e product page (Singapore)
- FDA decision summary K192240, RAPIDPoint 500e
- FDA decision summary K113216, RAPIDPoint 500 lactate
- Royal Children's Hospital Melbourne, RAPIDPoint 500e operator education sheet BI-W-116B
- openFDA recall search, RAPIDPoint 500 and 500e
- openFDA MAUDE problem counts, RAPIDPoint 500 family
Not covered: A copy of the RAPIDPoint 500e Operator's Guide is hosted by a Singapore hospital group but sat behind a bot check and could not be opened, so its error code tables, maintenance schedule, dimensions and power data are not included; only messages named in FDA records are listed. Cleaning and sample handling checks come from one hospital's training sheet. There are 18 recall records for the RAPIDPoint 500 family; older 500 only items (patient ID and data stream issues, 2012 to 2018) are left out. MAUDE counts mix the 500 and 500e and are trends only.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.