Model note · Blood Gas Analyser
Radiometer ABL800 FLEX blood gas analyser
The ABL800 FLEX is a bench top blood gas analyser used in intensive care, theatres, emergency departments and central labs to measure blood gases, electrolytes, metabolites and co-oximetry on one whole blood sample. It combines electrochemical sensors with a 128 wavelength spectrophotometer, runs automated quality control through its AutoCheck module, and with the FLEXQ module can scan, mix and analyse up to three barcoded syringes without staff standing over it.
The model
Maker, model and type.
- Maker
- Radiometer Medical ApS, Brønshøj, Denmark
- Model
- ABL800 FLEX blood gas analyser (models 393-800 and 393-801), optional FLEXQ autosampler
- Device type
- Blood Gas Analyser
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Parameters | Up to 18: pH, pCO2, pO2, cNa+, cK+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHHb, FHbF | Source |
| Minimum sample | 35 uL in FLEXMODE and MICROMODE | Source |
| Measuring technology | Electrochemical sensors for blood gases, pH, electrolytes, glucose and lactate; 128 wavelength spectrophotometer for co-oximetry and bilirubin | Source |
| FLEXMODE | Automatically measures fewer parameters when the sample is too small for the full panel | Source |
| Quality control | AutoCheck module for automated analysis of QC fluids | Source |
| FLEXQ module | Holds up to three safePICO samplers, reads their barcodes and mixes each by rotating magnet for about 5 seconds before measuring | Source |
| Other sample type | Expired air for pO2 and pCO2 | Source |
| FDA clearance | 510(k) K041874, decision 8 October 2004, 21 CFR 862.1120, class II; FLEXQ added under K043218 | Source |
Recalls
Known FDA recalls.
- Z-0038-20082007-06-22
Pump waste tubes made incorrectly could break at the connectors.
FDA record - Z-2571-20112011-04-07
Calcium membrane units could wrinkle and cause positive calcium bias.
FDA record - Z-1293-20152014-10-29
Negative QC drift during in use life of Crea A and B membranes.
FDA record - Z-0231-20182017-10-24
FLEXQ barcode scanner misreading a registration receipt could attach another patient's result.
FDA record - Z-0801-20192018-07-02
Creatinine could be underestimated by up to 28 percent.
FDA record - Z-1836-20202020-03-06
Incorrect power supply component could short circuit.
FDA record - Z-1684-20202020-03-10
Barcode reader misreading barcodes without a check digit.
FDA record - Z-2493-20202020-05-26
Potential patient mix up due to software issues.
FDA record - Z-1321-20222022-05-18
Sporadic biases for cNa+, cCa2+ and cK+.
FDA record - Z-2426-20242024-05-31
Falling calibration solution pH from bacterial growth could bias patient pH.
FDA record
Safety
Safety warnings from the record.
- Biased pH from degraded calibration solution can lead to wrong treatment decisions if not caught by QC. Source
- Creatinine could read up to 28 percent low against a reference method in some situations, a concern before contrast imaging. Source
- Several software and barcode recalls involved results being linked to the wrong patient; check patient identity on every report. Source
- Sporadic positive and negative biases were reported for sodium, calcium and potassium in some micro measuring modes on affected software. Source
Sources
Where this note comes from.
- Radiometer ABL800 FLEX product page
- FDA 510(k) summary K041874, ABL800 FLEX
- FDA 510(k) summary K043218, ABL800 FLEX with FLEXQ
- openFDA recall records, ABL800
- openFDA MAUDE problem counts, ABL800
Not covered: The ABL800 FLEX Operator's Manual (with its troubleshooting messages chapter) is not published openly by Radiometer, so no message codes, measuring times or reagent life figures are included; reseller figures for sample volume and cycle time were left out.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.