Radiometer ABL800 FLEX blood gas analyser

The ABL800 FLEX is a bench top blood gas analyser used in intensive care, theatres, emergency departments and central labs to measure blood gases, electrolytes, metabolites and co-oximetry on one whole blood sample. It combines electrochemical sensors with a 128 wavelength spectrophotometer, runs automated quality control through its AutoCheck module, and with the FLEXQ module can scan, mix and analyse up to three barcoded syringes without staff standing over it.

The model

Maker, model and type.

Maker
Radiometer Medical ApS, Brønshøj, Denmark
Model
ABL800 FLEX blood gas analyser (models 393-800 and 393-801), optional FLEXQ autosampler
Device type
Blood Gas Analyser
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ParametersUp to 18: pH, pCO2, pO2, cNa+, cK+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHHb, FHbFSource
Minimum sample35 uL in FLEXMODE and MICROMODESource
Measuring technologyElectrochemical sensors for blood gases, pH, electrolytes, glucose and lactate; 128 wavelength spectrophotometer for co-oximetry and bilirubinSource
FLEXMODEAutomatically measures fewer parameters when the sample is too small for the full panelSource
Quality controlAutoCheck module for automated analysis of QC fluidsSource
FLEXQ moduleHolds up to three safePICO samplers, reads their barcodes and mixes each by rotating magnet for about 5 seconds before measuringSource
Other sample typeExpired air for pO2 and pCO2Source
FDA clearance510(k) K041874, decision 8 October 2004, 21 CFR 862.1120, class II; FLEXQ added under K043218Source

Recalls

Known FDA recalls.

  • Z-0038-20082007-06-22

    Pump waste tubes made incorrectly could break at the connectors.

    FDA record
  • Z-2571-20112011-04-07

    Calcium membrane units could wrinkle and cause positive calcium bias.

    FDA record
  • Z-1293-20152014-10-29

    Negative QC drift during in use life of Crea A and B membranes.

    FDA record
  • Z-0231-20182017-10-24

    FLEXQ barcode scanner misreading a registration receipt could attach another patient's result.

    FDA record
  • Z-0801-20192018-07-02

    Creatinine could be underestimated by up to 28 percent.

    FDA record
  • Z-1836-20202020-03-06

    Incorrect power supply component could short circuit.

    FDA record
  • Z-1684-20202020-03-10

    Barcode reader misreading barcodes without a check digit.

    FDA record
  • Z-2493-20202020-05-26

    Potential patient mix up due to software issues.

    FDA record
  • Z-1321-20222022-05-18

    Sporadic biases for cNa+, cCa2+ and cK+.

    FDA record
  • Z-2426-20242024-05-31

    Falling calibration solution pH from bacterial growth could bias patient pH.

    FDA record

Safety

Safety warnings from the record.

  • Biased pH from degraded calibration solution can lead to wrong treatment decisions if not caught by QC. Source
  • Creatinine could read up to 28 percent low against a reference method in some situations, a concern before contrast imaging. Source
  • Several software and barcode recalls involved results being linked to the wrong patient; check patient identity on every report. Source
  • Sporadic positive and negative biases were reported for sodium, calcium and potassium in some micro measuring modes on affected software. Source