Model note · Transcranial Doppler
Rimed Digi-Lite all in one transcranial Doppler system
Digi-Lite is Rimed's cart based all in one TCD system for neurovascular labs, stroke units and intensive care. A PC platform with a touchscreen hosts a dedicated Doppler card with FPGA and DSP processing, driving 2 MHz pulsed probes for transcranial work and 4, 8 and 16 MHz probes for other vessels. It supports unilateral and bilateral monitoring, emboli detection, vasomotor reactivity tests and an LMY-3 headset with automatic signal tracking.
The model
Maker, model and type.
- Maker
- Rimed Ltd.
- Model
- Digi-Lite all in one transcranial Doppler system
- Device type
- Transcranial Doppler
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Doppler processing | One Doppler card with analogue and digital circuits, an FPGA for timing and a DSP for control and signal processing | Source |
| Gates and spectra | 64 gates (8000 channels); up to 8 spectral windows in diagnostic mode, 2 in monitoring | Source |
| Probes | 2 MHz PW, 4 MHz CW/PW, 8 MHz CW/PW, 16 MHz PW (16 MHz not cleared for the US); 5 probe connectors | Source |
| Display | 15 inch colour TFT LED touchscreen | Source |
| Computer | Windows 10, 4 GB RAM, 1 TB disk, 3 USB ports and LAN; 8 analogue inputs and outputs | Source |
| Electrical safety | Class 1, Type BF, EN 60601-1 and 60601-1-2; 100 to 240 V AC, 50/60 Hz | Source |
| Size and weight | 40 x 32 x 33 cm; 7 kg | Source |
| FDA clearance | K062578, product codes IYN and ITX, September 2006 | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Sources
Where this note comes from.
- FDA 510(k) K062578 summary, Digi-Lite TCD Device
- FDA 510(k) K143476 summary, Digi-One System
- Rimed product line catalogue (January 2022)
- Rimed Digi-Lite product page
Not covered: The 2006 510(k) describes the original Digi-Lite on Windows XP embedded; the catalogue describes a later Windows 10 build, and the Digi-Lite 2.0 brochure was not read. No operating or service manual was found, so error messages, calibration and service intervals are missing. No FDA recall or MAUDE records were found under Rimed or Digi-Lite.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Transcranial Doppler
Other models of this device.