Skip to content

Model note · Transcranial Doppler

Rimed Digi-Lite all in one transcranial Doppler system

Digi-Lite is Rimed's cart based all in one TCD system for neurovascular labs, stroke units and intensive care. A PC platform with a touchscreen hosts a dedicated Doppler card with FPGA and DSP processing, driving 2 MHz pulsed probes for transcranial work and 4, 8 and 16 MHz probes for other vessels. It supports unilateral and bilateral monitoring, emboli detection, vasomotor reactivity tests and an LMY-3 headset with automatic signal tracking.

The model

Maker, model and type.

Maker
Rimed Ltd.
Model
Digi-Lite all in one transcranial Doppler system
Device type
Transcranial Doppler
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Doppler processingOne Doppler card with analogue and digital circuits, an FPGA for timing and a DSP for control and signal processingSource
Gates and spectra64 gates (8000 channels); up to 8 spectral windows in diagnostic mode, 2 in monitoringSource
Probes2 MHz PW, 4 MHz CW/PW, 8 MHz CW/PW, 16 MHz PW (16 MHz not cleared for the US); 5 probe connectorsSource
Display15 inch colour TFT LED touchscreenSource
ComputerWindows 10, 4 GB RAM, 1 TB disk, 3 USB ports and LAN; 8 analogue inputs and outputsSource
Electrical safetyClass 1, Type BF, EN 60601-1 and 60601-1-2; 100 to 240 V AC, 50/60 HzSource
Size and weight40 x 32 x 33 cm; 7 kgSource
FDA clearanceK062578, product codes IYN and ITX, September 2006Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Not intended for fetal or neonatal use. Source
  • The 16 MHz intraoperative probe is not cleared for use in the US. Source

Sources

Where this note comes from.

Not covered: The 2006 510(k) describes the original Digi-Lite on Windows XP embedded; the catalogue describes a later Windows 10 build, and the Digi-Lite 2.0 brochure was not read. No operating or service manual was found, so error messages, calibration and service intervals are missing. No FDA recall or MAUDE records were found under Rimed or Digi-Lite.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.