Model note · Transcranial Doppler
DWL Doppler-BoxX transcranial Doppler unit
The Doppler-BoxX is Compumedics DWL's compact transcranial Doppler unit that plugs into an external Windows PC instead of carrying its own screen. It runs pulsed wave TCD with Doppler M-mode at 1, 2 and 2+2.5 MHz, CW and PW at 4 and 8 MHz, and a 16 MHz intraoperative microprobe, and supports routine exams, monitoring, emboli detection and CO2/VMR testing through software modules. It descends from the original Doppler-Box cleared by FDA in 2005.
The model
Maker, model and type.
- Maker
- Compumedics Germany GmbH (DWL)
- Model
- Doppler-BoxX transcranial Doppler unit (PC connected)
- Device type
- Transcranial Doppler
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Size and weight | 10 x 18.5 x 28 cm; 2.6 kg | Source |
| Doppler frequencies | PW 1, 2, 2+2.5 and 16 MHz; PW and CW 4 and 8 MHz (some not available in the USA) | Source |
| Channels and spectra | 1 to 2 Doppler channels; 1 to 9 spectrum windows | Source |
| Depth gates | Up to 400 gates and 8000 microgates | Source |
| Maximum measurable velocity | Up to 1900 cm/s on the data sheet (product page states 1,200 cm/s) | Source |
| Interfaces | 8 analogue inputs, 4 analogue outputs, Ethernet 10/100/1000; DICOM, HL7, GDT optional | Source |
| Monitor data links | Philips IntelliBridge and ICM+ (monitoring software only) | Source |
| Minimum PC operating system | Windows 10 Pro | Source |
| Regulatory | MDD Class IIa for system and probes; FDA K051085 | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Sources
Where this note comes from.
- DWL Doppler-BoxX data sheet Rev 12, 2023
- DWL Doppler-BoxX product page
- FDA 510(k) K051085, Doppler-Box
- DWL recommended detergents for cleaning and disinfection (PR-003)
Not covered: No operating or service manual was opened, so error messages, acoustic output calibration and maintenance intervals are missing. The 510(k) summary text is partly unreadable in the scanned PDF and covers the older Doppler-Box. The data sheet and product page disagree on maximum velocity (1900 vs 1,200 cm/s). No FDA recall or MAUDE records were found for this brand.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.