Natus Sonara transcranial Doppler system

The Sonara is a transcranial and peripheral vascular Doppler system that measures blood flow velocities in intracranial, extracranial and peripheral vessels without needles or contrast. It is used in neurology, stroke units, neurocritical care and vascular labs, including long term monitoring with a headset mounted probe. It comes either as an all in one cart with an integrated touchscreen PC or as Sonara/tek, the same hardware run from a separate computer over USB.

The model

Maker, model and type.

Maker
Natus Medical Incorporated (originally Viasys Healthcare NeuroCare)
Model
Sonara transcranial Doppler system (unilateral and bilateral; Sonara/tek PC version)
Device type
Transcranial Doppler
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Display and computer (Sonara)Integrated 15 inch colour touchscreen LCD with built in PC board and hard diskSource
Sonara/tek variantSame device operated from a personal computer through a USB connectionSource
Probe frequencies2 MHz, 4 MHz and 8 MHz, unilateral or bilateralSource
Monitoring probeDedicated 2 MHz monitoring probe that fits a headsetSource
DisplaysSpectral Doppler plus M-mode; output power shown at all times while scanningSource
AccessoriesRemote control and footswitch; online and offline operating modesSource
StandardsIEC/EN 60601-1, 60601-1-2 and 60601-2-37Source
FDA clearance510(k) K060421, decision 3 November 2006, class II, product codes IYN and ITXSource

Recalls

Known FDA recalls.

  • Z-0998-20092009-01-06

    Sonara unilateral: lock ups, shutdowns or system errors, and possible cooling fan wiring fault.

    FDA record
  • Z-0999-20092009-01-06

    Sonara bilateral: same lock up, shutdown and fan wiring issues.

    FDA record

Safety

Safety warnings from the record.

  • A staff member reported a shock when touching a 2 MHz probe to the neck; the unit and probe were returned for engineering inspection. Treat any such report as a removal from use event. Source
  • Acoustic output is kept below pre-amendment intensity levels and the power level is shown during scanning, so operators should keep exposure as low as practical. Source

Sources

Where this note comes from.

Not covered: The Natus product page blocked access and the Sonara user manual (Rev 09, 2016) is only available behind a terms agreement on a third party library, so no current specifications, troubleshooting steps or error messages could be checked; MAUDE holds only three reports.