Philips Ingenia 1.5T

The Philips Ingenia 1.5T is a 70 cm wide-bore superconducting MR scanner used for general clinical imaging in hospitals and imaging centres. Its distinctive feature is dStream, where the analogue-to-digital converter sits inside the receive coil so the signal travels digitally to the reconstructor. It also uses Philips HeliumSave technology, rated for zero helium boil-off during normal scanning.

The model

Maker, model and type.

Maker
Philips Medical Systems Nederland B.V.
Model
Ingenia 1.5T
Device type
MRI
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Field strength1.5 TSource
Patient bore70 cm, flared to a 95 cm tunnel diameter at both endsSource
Gradients (Omega HP)45 mT/m maximum amplitude and 200 T/m/s maximum slew rate, per axisSource
Magnet homogeneity0.9 ppm or better, typical, at 45 cm DSVSource
Maximum field of view55 cmSource
RF transmitParallel RF transmission, 18 kW output powerSource
Magnet weight3060 kg (total installed gantry weight 4600 kg or less)Source
Maximum patient weight250 kgSource
Minimum siting area27 m2Source
Cryogen boil-off0 l/hr during regular scanning (HeliumSave)Source
US regulatory statusClass II, regulation 21 CFR 892.1000, product code LNH; example clearance K173079 for software R5.4Source

Recalls

Known FDA recalls.

  • Z-2204-20252025-06-30

    Gradient coil component failures may act as a heat source and produce smoke or fire.

    FDA record
  • Z-1845-20242024-05-03

    g-MDU L3 terminal connection may loosen and cause a hot spot, smoke or fire.

    FDA record
  • Z-1829-20242024-04-22

    Ingenia Evolution systems on R5.7.1 may not warn about a low bore ventilation setting.

    FDA record
  • Z-0950-20242024-01-03

    Ingenia 1.5T CX Quadrature Body Coil seal adhesive may fail and leave sharp edges.

    FDA record
  • Z-2341-20212021-07-30

    Wrong pressure relief valve in the magnet assembly could release helium gas.

    FDA record
  • Z-0293-20122011-08-03

    RF heating could redden the skin of some larger patients; corrected by software.

    FDA record

Safety

Safety warnings from the record.

  • Gradient coil failures can produce smoke or fire. Philips advises stopping all scanning, including phantom scans, until service releases an affected system showing the warning signs. Source
  • In a developing fire the hospital procedure may include switching off the whole system or using the Emergency Magnet Off button to remove the field. Source
  • RF heating has caused skin reddening in some larger patients when the torso lies inside the body coil. Source
  • A trolley accessory was supplied with magnetic wheels that can be pulled towards the magnet, a reminder to check every item entering the scan room. Source
  • A wrongly rated pressure relief valve fitted to some magnets could release helium gas and cause injury. Source

Sources

Where this note comes from.

Not covered: No public Ingenia service manual or error code list was found; the only messages listed are those quoted in a Philips correction notice. MAUDE brand-name counts include coils and accessories, not only the scanner.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.