Model note · SPECT

Philips BrightView SPECT

BrightView is a Philips dual-detector gamma camera for SPECT and planar nuclear medicine, built on an open ring gantry with a large bore. It is used for cardiac, bone, renal and other single-photon studies, and the XCT variant adds a flat-panel CT for attenuation correction and localisation. Its distinctive points are the 9-axis ring gantry, automatic body contouring with BodyGuard sensing, and a cart-based collimator exchanger.

The model

Maker, model and type.

Maker
Philips Medical Systems (Cleveland), Inc.
Model
BrightView SPECT (product code 882480; BrightView X and XCT variants)
Device type
SPECT
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
DetectorsTwo detectors, usable singly or at relative 90 or 180 degreesSource
Detector field of view15.5 x 21 inch, square-cornered detectorsSource
Patient bore38 inch diameterSource
Gantry9-axis ring gantrySource
Brain reachUnder 3 inchesSource
Caudal and cephalic tiltAt least plus or minus 15 degrees on at least one detectorSource
Acquisition modesStatic, dynamic, gated, total body, circular and non-circular SPECT, gated SPECT and coincidence studiesSource
FDA clearance cited for BrightViewK062298 (Apollo Gamma Camera System), product code KPS, 21 CFR 892.1200, Class II, decision 23 August 2006; BrightView VCT cleared as K080927Source

Recalls

Known FDA recalls.

  • Z-0929-20242023-12-15

    Detector may fall unexpectedly due to a component failure.

    FDA record
  • Z-0677-20242023-11-30

    Pre-programmed motion in extrinsic QA could create a limb entrapment gap.

    FDA record
  • Z-2009-20192019-06-10

    Hand controller could cause uncommanded motion or motion after release.

    FDA record
  • Z-0510-20182017-12-15

    Collimator could drop during exchange if cart and detectors were misaligned.

    FDA record
  • Z-1750-20172017-02-24

    Brake hub not engaging gearbox shaft after service, so detector could backdrive.

    FDA record
  • Z-2735-20162016-05-25

    Software defect let a detector contact the patient in some non-ABC scans.

    FDA record

Safety

Safety warnings from the record.

  • A falling detector below the gantry centre could cause abrasion, contusion, laceration or fracture of a patient's lower limbs. Source
  • Pre-programmed motion during an extrinsic QA scan could open a gap between the patient support and detector, a limb entrapment risk. Source
  • A collimator can drop to the floor if the exchange cart and detectors are misaligned. Keep the area beside the cart clear during exchanges. Source
  • Gaps between the two parts of the patient table can pinch skin or fingers. Source

Sources

Where this note comes from.

Not covered: Philips no longer hosts a BrightView product page or datasheet, and no public operator or service manual or error code list was found. Specifications come from the 2006 510(k) summary for the Apollo system that FDA recall records cite for BrightView, so crystal thickness and current configuration details should be checked.