Model note · PCNL System
Olympus ShockPulse-SE Lithotripsy System
ShockPulse-SE is an intracorporeal lithotripter for urinary stones in the kidney, ureter and bladder, used through an endoscope. A reusable transducer converts electrical energy into ultrasonic vibration that travels down a probe to break up stones, and the system aspirates stone debris. It comprises a generator (SPL-G), transducer (SPL-T) and single use or reusable probes, with an optional footswitch.
The model
Maker, model and type.
- Maker
- Olympus (manufactured by Cybersonics, Inc.)
- Model
- ShockPulse-SE Lithotripsy System
- Device type
- PCNL System
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Components | Generator, probes, transducer, cleaning stylet, wrench, power cord, nosecone; optional footswitch | Source |
| Principle | Transducer converts electrical energy into ultrasonic vibration that travels down the probe; stone debris is aspirated | Source |
| Output and frequency | Maximum 100 W ultrasonic output to the transducer; 19,500 to 21,500 Hz | Source |
| Probes | 3.76, 3.40, 1.83, 1.50 and 0.97 mm; 304 stainless steel (larger sizes with diamond like carbon coating); single use EtO sterile or reusable steam sterilised (prevac) | Source |
| Model numbers | Generator SPL-G; transducer SPL-T; probes SPL-S (single use) and SPL-SR (reusable) | Source |
| Applied part and indications | Type BF; fragmentation of urinary tract calculi in the kidney, ureter and bladder | Source |
| FDA clearance | K171024 (August 2017), predicate K142428; product code FEO | Source |
Recalls
Known FDA recalls.
- Z-1494-20262026-01-08
Mis wired component adds power supply noise outside EMC standards (further serial numbers).
FDA record - Z-1479-20262025-11-14
Generator may stay blinking without recognising the transducer; plug or receptacle damage (also Z-1477 and Z-1478-2026 for probes).
FDA record - Z-0452-20262025-10-08
Transducer may fail to start or stop with an error light, and the handpiece may warm in use.
FDA record - Z-1514-20222022-07-07
Mis wired generator component causing power supply noise.
FDA record - Z-2197-20202020-04-23
Transducer component of non conforming material could allow water ingress during autoclaving.
FDA record
Sources
Where this note comes from.
- FDA 510(k) K171024 summary, ShockPulse-SE Lithotripsy System
- openFDA recall search, ShockPulse
- openFDA MAUDE problem counts, ShockPulse brand
Not covered: The Olympus IFU was not opened, so front panel indicators, power settings, reprocessing limits and probe life are missing. Recall records were read through openFDA. MAUDE counts may include bulk reports.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.