KARL STORZ Percutaneous Nephroscope System

The KARL STORZ Percutaneous Nephroscope System is a family of rigid rod lens nephroscopes, sheaths, dilators and hand instruments used in urology theatres to remove kidney stones through a track in the back (PCNL). Its MIP (minimally invasive percutaneous) range shrinks the instruments, for example a 12 Fr nephroscope in a 16.5/17.5 Fr sheath for MIP-M, and relies on controlled irrigation that flushes fragments out without a basket.

The model

Maker, model and type.

Maker
KARL STORZ SE & Co. KG (US submitter: KARL STORZ Endoscopy-America, Inc.)
Model
Percutaneous Nephroscope System (MIP nephroscopes and HOPKINS PCNL telescopes)
Device type
PCNL System
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
System contentsMIP nephroscopes, HOPKINS telescopes for PCNL, operating sheaths, one-step dilators, telescope bougie set, hollow obturator and fascial dilator, forceps, scissors, knives, instrument port, laser hand instrument, suction tube, adaptorsSource
OpticsRigid rod lens telescopesSource
MIP-M set12 Fr nephroscope, 16.5/17.5 Fr operating nephrostomy (Amplatz) sheath, single-step dilator and grasping forcepsSource
MIP-M stone sizeSuited to stones 12 to 24 mm in any calyx or the renal pelvisSource
Compatible lithotripsy (MIP-M)Lithotripsy devices down to 5 Fr, ballistic probes up to 2 mm, laser fibres 200 to 600 micrometresSource
Fascial dilator sizes9 to 30 Fr in 3 Fr stepsSource
IndicationsPercutaneous kidney stone removal in adults, and in children over 1 month for the paediatric configurationSource
Regulatory classFDA Class II nephroscope kit, 21 CFR 876.1500, product code FGASource
Standards testedISO 8600 endoscope series and IEC 60601-2-18 thermal safetySource

Recalls

Known FDA recalls.

  • Z-3224-20242024-07-23

    Nephroscope for MIP M (REF 27830KA): IFU contained reprocessing methods not reviewed or approved by the FDA.

    FDA record
  • Z-3225-20242024-07-23

    Nephroscope for MIP L (REF 27840KA): same reprocessing IFU issue.

    FDA record

Safety

Safety warnings from the record.

  • Some KARL STORZ IFUs, including those for MIP M and MIP L nephroscopes, listed reprocessing methods not reviewed by the FDA. Follow the corrected instructions. Source
  • Irrigation is run under controlled pressure, and the sheath design helps limit fluid overload; monitor irrigation pressure and volume throughout. Source
  • Using MIP-M for stones larger than its suggested range can lengthen surgery and raise complication risk. Source
  • Stone fragmentation equipment must be on hand and confirmed compatible with the MIP-M system before the case starts. Source

Sources

Where this note comes from.

Not covered: This is a passive optical instrument set with no error codes; no public KARL STORZ instructions for use or reprocessing manual were reviewed, so exact scope lengths, channel sizes per model and the repair inspection procedure are not included, and the first checks on lens and shaft inspection are general practice linked to MAUDE trends rather than a Storz procedure.