Model note · PCNL System
EMS Swiss LithoClast Trilogy Lithotripsy System
The LithoClast Trilogy is an intracorporeal lithotripter for kidney, ureter and bladder stones that combines ultrasonic vibration and electromagnetically generated ballistic impacts in one probe and one handpiece. A console sets the energy, a two step foot pedal controls suction and energy, and a peristaltic pump or pinch valve version clears fragments into an optional stone catcher.
The model
Maker, model and type.
- Maker
- EMS Electro Medical Systems SA (distributed by Boston Scientific)
- Model
- Swiss LithoClast Trilogy Lithotripsy System
- Device type
- PCNL System
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Operating modes | Ultrasound alone, ballistic alone, or combined | Source |
| Ultrasound | Transducer frequency 24 kHz (as listed for the predicate; same for Trilogy); maximum amplitude 60 um peak to peak; power 10 to 100% in 10% steps | Source |
| Ballistic | Generated electromagnetically (the predicate used compressed air); 10 to 100% in 10% steps; multiple shot mode only | Source |
| Probes | 1.1, 1.5, 1.9, 3.4 and 3.9 mm; one probe delivers both energies; RFID tag lets the console track probe type and use; EtO sterile, reusable probes validated for 4 steam cycles after first use | Source |
| Console | Touchscreen; 135 x 360 x 420 mm; 13.5 kg; 100 to 240 VAC, 50 to 60 Hz | Source |
| Handpiece | Single reusable PEEK handpiece; pre vacuum steam sterilisation | Source |
| Suction | Peristaltic pump or pinch valve model; pinch valve version needs an external vacuum | Source |
| FDA clearance | K173234 (January 2018), predicate LithoClast Master (K012445); later K181364, K181997, K191124, K230893; product code FEO | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Sources
Where this note comes from.
- FDA 510(k) K173234 summary, Swiss LithoClast Trilogy
- openFDA 510(k) search, LithoClast
- openFDA recall search, EMS Electro Medical Systems
- openFDA MAUDE problem counts, Swiss LithoClast Trilogy
Not covered: No FDA recall names the Trilogy; the two listed EMS lithotripsy recalls concern the older Master and Ultra and are used only as context. The Trilogy IFU was not opened, so error messages, handpiece reprocessing and power settings guidance are missing. The summary table prints the Trilogy ultrasound frequency as '24 Hz' marked 'same' as the predicate's 24 kHz, so 24 kHz is assumed.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
PCNL System