Olympus Endocapsule 10 System

The Endocapsule 10 System is Olympus's small intestine capsule endoscope. A single use capsule passes by peristalsis and transmits images to an eight element antenna sheet and a recorder worn by the patient; the recorder then docks in a cradle to download into Endocapsule Software 10 on a workstation. In the US it is cleared for adults to look for bleeding sources, Crohn's lesions and causes of iron deficiency anaemia missed by standard endoscopy.

The model

Maker, model and type.

Maker
Olympus Medical Systems Corp.
Model
Endocapsule 10 System (EC-10 capsule, RE-10 recorder, Endocapsule Software 10)
Device type
MCCE
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Capsule size (EC-Y0006 design, carried into EC-10)11 mm outer diameter, 26 mm longSource
Illumination and opticsFour white LEDs around a compact objective lens under a clear top coverSource
Radio link315 MHz transmitter in the capsuleSource
Capsule powerTwo silver oxide batteries; mercury free cells from the EC-10 (K163069)Source
Antenna unitIntegrated antenna sheet with eight antenna elements, worn at the waist in an antenna holder (a lead type antenna was added later)Source
System components (K183053)Capsule set MAJ-2027, recorder set MAJ-2029 with RE-10 recorder, battery pack, antenna unit, cradle, holders, capsule activator MAJ-1478, software MAJ-2188 and MAJ-2189Source
Intended population (US)Adults onlySource
Software level of concernModerateSource
FDA clearance (current system)K183053, decided 5 March 2019, product code NEZ, 21 CFR 876.1300Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • The US clearance restricts use to adults, unlike some competing capsules cleared from age two. Source
  • The Red Lesion Detector only flags frames that may show bleeding, angioectasia, red spots, ulcers or erosions; every frame still needs physician review. Source

Sources

Where this note comes from.

Not covered: Endocapsule is a passive capsule, not magnetically controlled; it is included as a widely installed capsule system in this family. Capsule size, LEDs and the 315 MHz link come from the 2013 summary; K163069 says the EC-10 is otherwise identical, but frame rate, field of view and battery life were not found in a public document. No openFDA recall matched the Endocapsule. The instruction manual was not reached, so no recorder messages or error codes are given; the manufacturerUrl is the Olympus corporate site, not a product page.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.