Model note · MCCE

AnX NaviCam Capsule Endoscope System

NaviCam is a magnetically controlled capsule endoscopy (MCCE) system: the patient swallows a camera capsule, and an operator steers it around the stomach with an external magnet on a motorised platform. It is used in clinics and hospitals as a sedation-free way to see the stomach and duodenum. It was the first system in its FDA device type, created through a De Novo in 2020.

The model

Maker, model and type.

Maker
AnX Robotica Corp.
Model
NaviCam Capsule Endoscope System with NaviCam Stomach Capsule (and NaviCam Xpress Stomach System)
Device type
MCCE
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
System componentsIngestible capsule, data recorder (AKR-1) in a patient vest, locator (AKS-1) and controller (NaviEC-1000)Source
Capsule size12 mm outer diameter, 28 mm long (NaviCam Stomach Capsule AKEM-11SW)Source
Capsule imaging and powerCMOS sensor with LEDs and photoresistor around the lens; RF transmitter; two silver oxide batteriesSource
ControllerConsole, translational and rotational platform, magnetic ball and examination bedSource
Capsule control axesX, Y and Z translation plus clockwise and anticlockwise rotation, in real timeSource
Capsule radio frequency (Xpress, 2024)915 MHz, changed from 2.4 GHz for lower SARSource
Capsule shelf life14 months; supplied sterile (ethylene oxide)Source
Current indication (Xpress)Stomach visualisation in patients 6 years or older, BMI 65 or less, waist 77 inches or lessSource
FDA classificationDe Novo DEN190037, 21 CFR 876.1310, Class II, product code QKZSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • An MRI scan with the capsule still inside may damage the gut or abdomen; if excretion is not confirmed, the patient needs medical review and possibly an abdominal X-ray first. Source
  • The external magnet can interact with ferromagnetic implants in patients and users. Source
  • Capsule retention from obstruction, and aspiration of the capsule, are recognised risks. Source
  • False negative and false positive findings are the main clinical risks named by FDA. Source

Sources

Where this note comes from.

Not covered: The AnX Robotica website did not respond, and no public user or service manual was found, so frame rate, battery life, magnet strength and controller error messages are not included. There are no FDA recalls and very few MAUDE reports, so failing parts cannot be identified.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.