Nonin Onyx Vantage 9590 Finger Pulse Oximeter

The Onyx Vantage 9590 is a sealed fingertip pulse oximeter for spot checks of SpO2 and pulse rate on adults and children with fingers 0.8 to 2.5 cm thick. It runs on two AAA cells, switches on when a finger is inserted, and shows signal quality on a three colour LED. It has no alarms, is not repairable in the field and needs no calibration beyond battery changes.

The model

Maker, model and type.

Maker
Nonin Medical, Inc.
Model
Onyx Vantage 9590 Finger Pulse Oximeter
Device type
Pulse Oximeter
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Display rangesSpO2 0 to 100%; pulse rate 18 to 321 bpmSource
SpO2 accuracy (Arms, no motion)Finger plus or minus 2 digits at 70 to 100%; low perfusion plus or minus 2 digitsSource
Pulse rate accuracyPlus or minus 3 digits at 20 to 250 bpm; low perfusion plus or minus 3 digits at 40 to 240 bpmSource
Optical emittersRed 660 nm (0.8 mW max average); infrared 910 nm (1.2 mW max average)Source
BatteryTwo 1.5 V AAA alkaline cells, about 6,000 spot checks or 36 h continuous useSource
EnvironmentOperating minus 5 to 40 C, 10 to 90% RH, up to 4000 m; hyperbaric use to 4 atmospheresSource
ClassificationType BF applied part, IP32, continuous operation; ISO 80601-2-61; not defibrillation proofSource
Averaging and response4 beat exponential average with 2 beat latency for SpO2 and pulse rateSource
Expected service life5 yearsSource
FDA clearanceK112843, cleared 19 April 2012 (cited as predicate in K131021); product code DQASource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • No audible alarms: for spot checks only. Source
  • Do not use in MRI, in explosive atmospheres or on neonates. Source
  • Readings can be degraded by motion, bright light, nail polish, dyes, carboxy or methaemoglobin, anaemia and a cuff or line on the same arm. Source
  • Check the finger site at least every 4 hours if left on, and avoid extra pressure on the sensor. Source
  • Keep phones and radios at least 30 cm away, and expect interference from electrosurgery. Source

Sources

Where this note comes from.

Not covered: No FDA recall names the 9590; Nonin's four recalls cover other models. MAUDE counts filtered by model number 9590 appear to mix in unrelated devices (several are lab test result codes), so no MAUDE trend is given. The 9590's own 510(k) summary (K112843) was not opened; the clearance date is taken from a later Nonin summary. The 'housing seals and optics' entry reflects manual cautions, not failure data.