Medtronic Nellcor Bedside SpO2 Patient Monitoring System PM100N

The PM100N is a bedside pulse oximeter for continuous SpO2 and pulse rate monitoring of adults, children and neonates, including poorly perfused patients, in hospitals, hospital type facilities, transport and home care. It uses the same OxiMax oximetry board and algorithm as the earlier N-600x and works with standard Nellcor sensors. SatSeconds alarm management, a 96 hour trend memory and nurse call output are its main ward features.

The model

Maker, model and type.

Maker
Covidien LLC (Medtronic)
Model
Nellcor Bedside SpO2 Patient Monitoring System PM100N
Device type
Pulse Oximeter
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Measurement rangesSpO2 1 to 100%; pulse rate 20 to 250 bpm; pulse amplitude 0.03 to 20%Source
SpO2 accuracy (70 to 100%, plus or minus 1 SD)Adult and low perfusion plus or minus 2 digits; neonate plus or minus 3 digitsSource
Pulse rate accuracyPlus or minus 3 digits over 20 to 250 bpm (adult and low perfusion)Source
Trend memory96 hours at one sample every 4 seconds, downloadable to a PCSource
Power and battery100 to 240 VAC, 50/60 Hz, 45 VA; lithium ion battery, at least 8 h new with no alarms (8 h and 16 h packs offered)Source
Size, weight and ingress82 x 255 x 165 mm; 1.6 kg; IPX2Source
EnvironmentOperating 5 to 40 C, 15 to 93% RH non condensing, minus 170 to 4877 mSource
Alarm featuresHigh and low SpO2 and pulse rate, backup audible alarm, SatSeconds 10, 25, 50, 100 or off, low battery and sensor off warningsSource
Electrical classificationInternally powered, type BF applied part, continuous operationSource
FDA clearanceK120773 (July 2012), later K142865; product code DQA, Class II, 21 CFR 870.2700; predicate N-600x (K083325)Source

Recalls

Known FDA recalls.

  • Z-2151-20252025-06-23

    Labelling correction (264,790 units) after alarms were not heard or recognised; reiterates alarm volume, sensor off priority and Homecare Mode guidance.

    FDA record

Safety

Safety warnings from the record.

  • Missed alarms have been linked to treatment delay, unanswered low saturation, respiratory failure and arrhythmia. Source
  • Outside Homecare Mode, alarms can be silenced and limits changed, so a monitor sent home must be switched to Homecare Mode and the Home Use Guide given to carers. Source
  • Not suitable for use with flammable anaesthetics. Source

Sources

Where this note comes from.

Not covered: The operator's and service manuals were not reachable (the hosting library did not resolve and Medtronic's manuals portal was not opened), so the troubleshooting table and numbered error codes are missing; the single entry is the sensor off alert as described in the FDA recall record. The 'sensors' failing part is an interpretation of MAUDE reading codes, flagged as such. Recall records were read through openFDA.