Masimo Rad-8 Signal Extraction pulse oximeter

The Masimo Rad-8 is a compact mains and battery powered bedside pulse oximeter that continuously monitors SpO2, pulse rate and perfusion index in wards, recovery, neonatal units and home care. It uses Masimo SET signal extraction, designed to keep readings reliable during patient motion and low perfusion, and its internal battery runs it for up to 7 hours.

The model

Maker, model and type.

Maker
Masimo Corporation
Model
Rad-8 Signal Extraction pulse oximeter (bedside)
Device type
Pulse Oximeter
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
SpO2 range1 to 100 percent, 1 percent resolutionSource
Pulse rate range25 to 240 bpm, 1 bpm resolutionSource
Perfusion index range0.02 to 20 percentSource
SpO2 accuracy (70 to 100 percent)No motion: plus or minus 2 percent adults, infants, paediatrics, 3 percent neonates; motion: plus or minus 3 percentSource
Pulse rate accuracyPlus or minus 3 bpm no motion and low perfusion; plus or minus 5 bpm with motionSource
WavelengthsRed 660 nm and infrared 905 nmSource
Power100 to 240 V AC, 47 to 63 Hz, 20 VA maximumSource
BatterySealed lead acid, up to 7 hours, 8 hour chargeSource
Size and weight20.8 x 15.2 x 7.6 cm; about 0.9 kgSource
Trend memory72 hours at 2 second resolutionSource
Operating environment5 to 40 C; 500 to 1060 mbarSource

Recalls

Known FDA recalls.

  • Z-1187-20132013-03-13

    About 0.02 percent of Rad-8 units with a 20 pin cable connector could power off when moved.

    FDA record

Safety

Safety warnings from the record.

  • Raised carboxyhaemoglobin or methaemoglobin makes SpO2 inaccurate; high COHb can coexist with a normal looking SpO2, so use co-oximetry when suspected. Source
  • Not an apnoea monitor and not a substitute for ECG arrhythmia analysis. Source
  • Do not use during MRI; induced currents may cause burns. Source
  • Do not open the instrument because of electric shock risk; refer repairs to Masimo. Source
  • Unexpected power off removes monitoring without warning, which was the hazard behind the 2013 recall. Source

Sources

Where this note comes from.

Not covered: The Rad-8 service manual and its numeric system error codes were not reviewed from an official source, and no 510(k) summary was checked, so only the operator level display messages are listed.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.