Model note · Pulse Oximeter
Masimo Rad-8 Signal Extraction pulse oximeter
The Masimo Rad-8 is a compact mains and battery powered bedside pulse oximeter that continuously monitors SpO2, pulse rate and perfusion index in wards, recovery, neonatal units and home care. It uses Masimo SET signal extraction, designed to keep readings reliable during patient motion and low perfusion, and its internal battery runs it for up to 7 hours.
The model
Maker, model and type.
- Maker
- Masimo Corporation
- Model
- Rad-8 Signal Extraction pulse oximeter (bedside)
- Device type
- Pulse Oximeter
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| SpO2 range | 1 to 100 percent, 1 percent resolution | Source |
| Pulse rate range | 25 to 240 bpm, 1 bpm resolution | Source |
| Perfusion index range | 0.02 to 20 percent | Source |
| SpO2 accuracy (70 to 100 percent) | No motion: plus or minus 2 percent adults, infants, paediatrics, 3 percent neonates; motion: plus or minus 3 percent | Source |
| Pulse rate accuracy | Plus or minus 3 bpm no motion and low perfusion; plus or minus 5 bpm with motion | Source |
| Wavelengths | Red 660 nm and infrared 905 nm | Source |
| Power | 100 to 240 V AC, 47 to 63 Hz, 20 VA maximum | Source |
| Battery | Sealed lead acid, up to 7 hours, 8 hour charge | Source |
| Size and weight | 20.8 x 15.2 x 7.6 cm; about 0.9 kg | Source |
| Trend memory | 72 hours at 2 second resolution | Source |
| Operating environment | 5 to 40 C; 500 to 1060 mbar | Source |
Recalls
Known FDA recalls.
- Z-1187-20132013-03-13
About 0.02 percent of Rad-8 units with a 20 pin cable connector could power off when moved.
FDA record
Safety
Safety warnings from the record.
- Raised carboxyhaemoglobin or methaemoglobin makes SpO2 inaccurate; high COHb can coexist with a normal looking SpO2, so use co-oximetry when suspected. Source
- Not an apnoea monitor and not a substitute for ECG arrhythmia analysis. Source
- Do not use during MRI; induced currents may cause burns. Source
- Do not open the instrument because of electric shock risk; refer repairs to Masimo. Source
- Unexpected power off removes monitoring without warning, which was the hazard behind the 2013 recall. Source
Sources
Where this note comes from.
- Masimo Rad-8 operator's manual LAB-5303C
- Masimo addendum Rad-8 alarms and messages LAB-9101A
- openFDA recall Z-1187-2013
- openFDA MAUDE problem counts, Rad-8
- openFDA MAUDE narratives, Rad-8 power failures
Not covered: The Rad-8 service manual and its numeric system error codes were not reviewed from an official source, and no 510(k) summary was checked, so only the operator level display messages are listed.