Model note · Pacemaker

Medtronic Azure XT and Azure S MRI SureScan pacemakers

Azure is Medtronic's family of single and dual chamber MRI conditional pacemakers for bradycardia, implanted by cardiologists and followed up in device clinics. Its distinctive feature is BlueSync Bluetooth telemetry, which lets the implant talk to the tablet-based CareLink SmartSync programmer and to patient phone apps for remote monitoring, alongside AF detection, Reactive ATP and the MVP mode that limits unneeded right ventricular pacing.

The model

Maker, model and type.

Maker
Medtronic, Inc.
Model
Azure XT and Azure S MRI SureScan pacemakers (DR and SR)
Device type
Pacemaker
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ModelsAzure XT DR W1DR01, Azure XT SR W1SR01, Azure S DR W3DR01, Azure S SR W3SR01Source
ConnectorIS-1 BI/UNISource
Mass22.5 gSource
Volume12.75 mL (DR) and 12.25 mL (SR)Source
Size (H x W x D)46.6 x 50.8 x 7.4 mm (DR); 42.6 x 50.8 x 7.4 mm (SR)Source
TelemetryBlueSync Bluetooth, used with CareLink SmartSync device manager, MyCareLink Heart app or MyCareLink RelaySource
AF detection accuracy (manufacturer claim)95 to 96% episode positive predictive valueSource
PMAApproved under PMA P980035 supplements (for example S539, 2018)Source

Recalls

Known FDA recalls.

  • Z-2507-20202020-06-04

    Rare SmartSync first-interrogation sequence could temporarily suspend battery measurement, Capture Management and other features (Azure XT DR); sister entries cover XT SR, S DR and S SR.

    FDA record
  • Z-1525-20212021-04-06

    Longevity estimation error early in device life with a unipolar RA or RV pacing vector; fixed by software update.

    FDA record
  • Z-1924-20252025-04-29

    Azure S DR (W3DR01): possible sudden battery failure without alarm, power-on resets and interruption of therapy.

    FDA record
  • Z-2127-20252025-06-17

    SmartSync software update to remove a possibly erroneous electrical reset pop-up on interrogation.

    FDA record

Safety

Safety warnings from the record.

  • Some Azure S DR devices may lose battery power suddenly without an alarm, interrupting pacing; pacemaker-dependent patients are at risk of syncope or cardiac arrest. Source
  • MRI conditions are model and lead specific; check them with the Medtronic MR conditional tool rather than assuming any Azure system is safe to scan. Source
  • Electromagnetic interference is a recurring MAUDE problem term (about 410 reports) for Azure, so ask about EMI exposure when investigating unexpected behaviour. Source

Sources

Where this note comes from.

Not covered: The Azure device manual (M964338A001B) and the Medtronic product performance report could not be opened, so programmable ranges, longevity tables, RRT and EOS indicators and MRI scan conditions are not included; implants have no user error codes, and programmer messages are not listed.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.