Model note · Pacemaker

Abbott Assurity MRI pacemakers PM1272

Assurity MRI is Abbott's MR Conditional bradycardia pacemaker family, built on the Assurity platform with extra hardware to tolerate MRI scanning. It is programmed with the Merlin PCS programmer, followed remotely through Merlin.net, and paired with the Tendril MRI lead as part of an MR Conditional system. For hospital teams the main touch points are programmer software, MRI scan preparation and recall driven follow up rather than bench repair.

The model

Maker, model and type.

Maker
Abbott (St. Jude Medical)
Model
Assurity MRI pacemakers PM1272 (single chamber) and PM2272 (dual chamber)
Device type
Pacemaker
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ModelsAssurity MRI PM1272 and PM2272; sister Endurity MRI PM1172 and PM2172Source
MRI conditions (original approval)1.5 T, 64 MHz, cylindrical bore, spatial gradient 30 T/m, whole body SAR 2 W/kg, head SAR 3.2 W/kg, slew rate 200 T/m/s, full bodySource
MRI hardware changesLower feedthrough capacitance, band stop MR filter assembly, reshaped can, unique radiopaque MRI x ray markerSource
MRI Settings modesAsynchronous pacing or pacing off, with sensing disabled during the scanSource
Pulse amplitude tolerance (test requirement)5 V peak plus or minus 20%, from beginning of life to ERISource
Shelf life18 months, limited by the batterySource
Companion leadTendril MRI LPA1200M: 7.9 F, steroid eluting, bipolar, active fixation, IS-1Source
Support toolsMerlin PCS programmer software 3330, MRI Activator EX4000 handheldSource
FDA approvalPMA P140033, MR Conditional pacing systemSource

Recalls

Known FDA recalls.

  • Z-1530-20212021-03-15

    Class 1: moisture ingress through header and feedthrough from incomplete epoxy mixing (2015 to 2018 builds).

    FDA record
  • Z-0031-20182017-08-28

    Firmware update to reduce risk of unauthorised RF access (Assurity MRI PM1272, PM2272).

    FDA record

Safety

Safety warnings from the record.

  • Sensing is disabled in MRI Settings, so the patient is paced asynchronously or not at all during the scan. Source
  • Rate adaptive pacing may be unsuitable for patients with angina or other symptoms worsened by faster heart rates. Source
  • Moisture affected devices may reach ERI or EOS without warning; pacemaker dependent patients who cannot be remotely followed need extra risk review. Source

Sources

Where this note comes from.

Not covered: Battery capacity, projected longevity, size and weight are in Abbott's reference manual, which was not read. MRI conditions listed are from the 2014 approval; later supplements may have widened them (for example to 3 T), so the current MRI-Ready Systems Manual must be checked. Implanted devices have no user error codes. MAUDE counts include many lead and clinical reports and are not generator failure rates. The manufacturerUrl points to the FDA summary.