Model note · Laryngoscope
KARL STORZ C-MAC Video Laryngoscope
C-MAC is a modular video laryngoscope system: reusable Macintosh, Miller and hyperangulated D-BLADE blades with a CMOS camera and LEDs near the tip plug into a C-MAC monitor or pocket monitor. Because the standard blades are shaped like conventional Macintosh blades, it can be used for direct and video laryngoscopy and for teaching. Reprocessing of the reusable blades is a large part of its upkeep.
The model
Maker, model and type.
- Maker
- KARL STORZ SE & Co. KG
- Model
- C-MAC Video Laryngoscope (8401/8403 series blades with C-MAC monitor)
- Device type
- Laryngoscope
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| D-BLADE construction | Reusable, hyperangulated, non channelled titanium blade | Source |
| D-BLADE camera and light | CMOS camera, 1280 x 800 px; two 1 W white LEDs | Source |
| D-BLADE dimensions | Working length 109 mm; handle width 28 mm; distal tip width 19 mm | Source |
| Connection | Plug and play to C-MAC monitors (for example 8403 ZX or 8104 ZX) via a 200 cm connection cable | Source |
| Blade series in the 2019 recall | 8403xx (8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC) and 8401xx (8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC) | Source |
| 8403 ZX monitor status | Listed by a reseller as discontinued by the manufacturer; pole clamp and roll stand mounting | Source |
Recalls
Known FDA recalls.
- Z-1618-20202019-12-06
Distal glass cover over camera and LEDs could loosen during reprocessing or use; IFU revised and users retrained (27,165 scopes).
FDA record
Sources
Where this note comes from.
- openFDA recall search, C-MAC
- openFDA MAUDE problem counts, C-MAC brand
- LARINGOCOL trial protocol (ClinicalTrials.gov NCT07239141), D-BLADE description
- Ampronix listing, KARL STORZ C-MAC 8403ZX monitor
Not covered: This is the thinnest note in the batch. KARL STORZ IFUs are on its IFU portal and copies on a third party site refused access, so there are no monitor specs, error messages, reprocessing cycle limits or battery data. D-BLADE figures come from a clinical trial protocol, not KARL STORZ. No FDA 510(k) summary was found. A reviewer may prefer to replace this model with one that has a public manual.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Laryngoscope