Skip to content

Model note · Retinal Laser

IRIDEX IQ 532 Laser System

The IRIDEX IQ 532 is a compact 532 nm green photocoagulation laser used in eye clinics and theatres for retinal photocoagulation, trabeculoplasty and iridotomy, delivered through a slit lamp adapter, laser indirect ophthalmoscope or endoprobe. Besides conventional CW-Pulse it can deliver MicroPulse trains for subvisible threshold treatment, and its console identifies delivery devices and safety filters through a SmartKey and connector ports.

The model

Maker, model and type.

Maker
IRIDEX Corporation
Model
IQ 532 Laser System
Device type
Retinal Laser
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Wavelength532 nm treatment; 635 nm diode aiming beam, under 1 mWSource
Treatment power50 to 2500 mW delivered depending on delivery device; 50 to 5000 mW with XP optionSource
Exposure durationCW-Pulse 10 to 3000 ms or CW up to 60 s; MicroPulse 0.05 to 1.0 msSource
Repeat interval10 to 3000 ms or single pulse; MicroPulse interval 1.0 to 10.0 msSource
Electrical100 to 240 V AC, 50/60 Hz, under 3 ASource
CoolingAir cooledSource
Operating conditions10 to 35 C, 20 to 80% RH non-condensingSource
Size and weight30.5 x 35.6 x 21.4 cm, 9 kgSource
Wireless footswitchWorks within 5 m; battery life about 3 to 5 years; each footswitch and receiver pair is uniquely linkedSource
FDA clearanceFamily of IRIDEX IQ Laser Systems, K071687 (2008)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Everyone in the room must wear laser safety eyewear of the correct wavelength and optical density; prescription glasses are not a substitute. Source
  • Do not remove covers: there are no user-serviceable parts and opening it exposes dangerous laser radiation and voltages. Source
  • Keep the laser in Standby except during actual treatment. Source
  • Do not use near flammable anaesthetics, alcohol or skin prep solutions, and avoid aiming at reflective instruments. Source
  • Use only Iridex delivery devices, as others may deliver power inaccurately; connect only to earthed mains. Source

Sources

Where this note comes from.

Not covered: MAUDE holds only one report under this brand and no FDA recall names the IQ 532 console (the only IRIDEX recall found is for the TruFocus LIO Premiere), so common failing parts come from the operator manual rather than field data; the service manual is not public.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.