Model note · Retinal Laser
Ellex Integre Pro Scan pattern scanning photocoagulator
The Integre Pro Scan is an all in one slit lamp and laser photocoagulator, with the laser cavity built into the slit lamp table rather than a separate console. It offers single spot treatment and pattern scanning for panretinal photocoagulation, controlled from a tablet style interface. It comes in green 532 nm, yellow 561 nm or yellow and red configurations; this note focuses on the green version.
The model
Maker, model and type.
- Maker
- Ellex Medical Pty Ltd, now Lumibird Medical
- Model
- Integre Pro Scan pattern scanning photocoagulator (green 532 nm configuration, LP6G)
- Device type
- Retinal Laser
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Wavelength options | Green 532 nm; yellow 561 nm; yellow and red 561 and 670 nm | Source |
| Power at fibre port | Green 1.5 W (yellow 1.5 W, red 1 W) | Source |
| Power at cornea | Green 50 to 1500 mW | Source |
| Exposure time | Pattern scanning 10 to 30 ms; single spot 10 ms to 8.0 s | Source |
| Spot size | Pattern scanning 100 to 500 micrometres; single spot 50 to 1000, continuously variable | Source |
| Aiming beam | Red 635 nm, adjustable intensity, Class 2 below 1 mW | Source |
| Laser class | Class 4; Nd:DPSS 532 nm, 2 W maximum | Source |
| Electrical and cooling | 100 to 240 VAC, 50/60 Hz, 800 VA; air cooled | Source |
| Slit lamp magnification | 6x, 10x, 16x, 25x, 40x | Source |
| Weight and size | 35 kg (laser only); 62 x 76 x 47 cm | Source |
Recalls
Known FDA recalls.
- Z-1971-20232023-03-16
Integre Pro and Pro Scan models including Pro Scan Green LP6G: with an LIO fitted, scattered beam could exit through the slit lamp port.
FDA record
Sources
Where this note comes from.
- Lumibird Medical Integre Pro laser range brochure (8448344EN-A)
- FDA recall Z-1971-2023, Integre Pro family scattered beam
- FDA 510(k) K142398 decision letter, Integre Pro Scan
- openFDA MAUDE problem counts, brand Integre Pro Scan
Not covered: The FDA file for K142398 holds only the decision letter, not a technical summary. No operator or service manual was found, so error codes, calibration and maintenance intervals are missing. Brochure specifications vary slightly between Lumibird documents (for example green power at the fibre port), and the 2026 brochure was used.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Retinal Laser