GE Giraffe Incubator and Giraffe Incubator Carestation CS1

The Giraffe Incubator is GE's closed neonatal incubator, first cleared in 2001. The Carestation CS1 is its current version, alongside the Giraffe OmniBed, which can also open as a radiant warmer. It controls temperature by baby skin or by air, with servo humidity and an optional Servo Oxygen Module. In 2021 the porthole latches were redesigned to rotate to open, with a secondary catch. Its recalls are worth knowing for any neonatal unit: panels and portholes that look closed but are not latched (fall risk), oxygen that differs from the display on pre 2012 units, leakage current through the USB and nurse call ports, and formaldehyde from new units in their first week.

The model

Maker, model and type.

Maker
Datex-Ohmeda, Inc. (formerly Ohmeda Medical; GE HealthCare)
Model
Giraffe Incubator and Giraffe Incubator Carestation CS1
Device type
Infant Incubator
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classification and clearancesNeonatal incubator, product code FMZ; Giraffe Incubator from K010222 (2001); Incubator Carestation CS1 K152809 (2015), K213553 (2 March 2022, porthole latch redesign), K251663 (2025), K260326 (28 July 2026)Source
Control settings (CS1 family)Baby (patient) control 35 to 37.5 degrees C and air control 20 to 39 degrees C, both in 0.1 degree steps; servo humidity 30 to 95% relative humidity in 5% stepsSource
Operating environment20 to 30 degrees C room, 10 to 95% humidity non condensing, room air velocity up to 0.3 m/sSource
Latches (2021 design)Portholes rotate clockwise or anticlockwise to open; north side wall latches have a primary and a secondary latching point; side panels open by pinchingSource
OptionsServo Oxygen Module; USB and RS232 nurse call ports; uninterruptible power supply option (2007 clearance)Source

Recalls

Known FDA recalls.

  • Z-0141-20252023-06-06

    Higher airborne formaldehyde from newly made incubators in their first week (224 units).

    FDA record
  • Z-2236-20202020-04-28

    Pre 2012 units with the Servo Oxygen Module may deliver oxygen different from the display (5,883 units).

    FDA record
  • Z-0540-20202019-10-11

    Bedside panels and portholes can look closed without being latched; fall risk (3,467 units).

    FDA record
  • Z-0480-20192018-09-25

    Possible electric leakage affecting the user (1,724 devices).

    FDA record
  • Z-1502-20172016-11-18

    Certain power cords could overheat.

    FDA record
  • Z-1051-20132013-01-08

    Display boards may overwrite stored settings and set points at power up or after a power failure.

    FDA record

Safety

Safety warnings from the record.

  • An unlatched panel or porthole can let the baby fall. Source
  • Until tested, GE advised disconnecting devices from the USB and RS232 nurse call ports because of possible leakage current to the user. Source
  • Newly made units could release formaldehyde that falls quickly over about a week (224 units, 2023). Source

Sources

Where this note comes from.

Not covered: The comparison table in K213553 is shared with the Giraffe OmniBed Carestation, so some settings may be family values rather than incubator only; the radiant heat line was left out for that reason. The service manual and IFU were not read, so alarm messages, weigh scale accuracy, battery time and maintenance intervals are not included. The skin probe point in the design note is general incubator practice, not a GE statement.