Model note · Thermometer
Braun ThermoScan PRO 6000 Ear Thermometer
The ThermoScan PRO 6000 is a professional infrared tympanic thermometer for patients of all ages. It reads the infrared energy from the eardrum and nearby surfaces through a disposable probe cover, using a heated tip and ambient correction, and gives a result in 2 to 3 seconds. Its main change from the PRO 4000 is a technique compensation feature that uses three capacitive rings to check probe placement.
The model
Maker, model and type.
- Maker
- Kaz USA, Inc. (Helen of Troy); distributed by Welch Allyn / Hillrom (Baxter)
- Model
- ThermoScan PRO 6000 Ear Thermometer
- Device type
- Thermometer
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Measurement principle | Infrared detection with predictive algorithm and ambient temperature correction; heated probe tip | Source |
| Accuracy (laboratory) | Plus or minus 0.2 C from 35 to 42 C; plus or minus 0.3 C below 35 C and above 42 C | Source |
| Precision and repeatability | SD under 0.6 C; repeatability under 0.3 C | Source |
| Measurement range and response | 20 to 42.2 C; 2 to 3 seconds | Source |
| Operating environment | 10 to 40 C ambient, up to 95% RH | Source |
| Size and power | 150 x 60 x 35 mm; two AA alkaline cells or a custom NiMH rechargeable pack | Source |
| US models | PRO6000SCMN (small cradle), PRO6000LCMN (large cradle), PRO6000BSNA (charge dock) | Source |
| Standards and clearance | ASTM E1965, EN 12470-5, ISO 80601-2-56; K152748 (March 2016), predicate PRO 4000 (K103800); product code FLL, Class II | Source |
Recalls
Known FDA recalls.
- Z-2339-20242024-06-13
Some units shipped with a CD holding an outdated IFU (265,647 units).
FDA record - Z-2482-20212021-07-20
Fluid ingress and use before cleaning fluid dried could overheat the device and cause burns (1,036,740 units).
FDA record
Sources
Where this note comes from.
- FDA 510(k) K152748 summary, Braun ThermoScan PRO 6000
- openFDA recall search, PRO 6000
- openFDA MAUDE problem counts, ThermoScan PRO 6000
Not covered: The PRO 6000 IFU was not opened, so display messages (such as position or cover errors), cleaning agents and battery life are missing; the one indicator listed comes from a recall notice. MAUDE has only two reports for the brand name.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
Thermometer