Braun ThermoScan PRO 6000 Ear Thermometer

The ThermoScan PRO 6000 is a professional infrared tympanic thermometer for patients of all ages. It reads the infrared energy from the eardrum and nearby surfaces through a disposable probe cover, using a heated tip and ambient correction, and gives a result in 2 to 3 seconds. Its main change from the PRO 4000 is a technique compensation feature that uses three capacitive rings to check probe placement.

The model

Maker, model and type.

Maker
Kaz USA, Inc. (Helen of Troy); distributed by Welch Allyn / Hillrom (Baxter)
Model
ThermoScan PRO 6000 Ear Thermometer
Device type
Thermometer
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Measurement principleInfrared detection with predictive algorithm and ambient temperature correction; heated probe tipSource
Accuracy (laboratory)Plus or minus 0.2 C from 35 to 42 C; plus or minus 0.3 C below 35 C and above 42 CSource
Precision and repeatabilitySD under 0.6 C; repeatability under 0.3 CSource
Measurement range and response20 to 42.2 C; 2 to 3 secondsSource
Operating environment10 to 40 C ambient, up to 95% RHSource
Size and power150 x 60 x 35 mm; two AA alkaline cells or a custom NiMH rechargeable packSource
US modelsPRO6000SCMN (small cradle), PRO6000LCMN (large cradle), PRO6000BSNA (charge dock)Source
Standards and clearanceASTM E1965, EN 12470-5, ISO 80601-2-56; K152748 (March 2016), predicate PRO 4000 (K103800); product code FLL, Class IISource

Recalls

Known FDA recalls.

  • Z-2339-20242024-06-13

    Some units shipped with a CD holding an outdated IFU (265,647 units).

    FDA record
  • Z-2482-20212021-07-20

    Fluid ingress and use before cleaning fluid dried could overheat the device and cause burns (1,036,740 units).

    FDA record

Safety

Safety warnings from the record.

  • Fluid ingress followed by use before drying can overheat the device and burn the user's hand or the patient's ear canal. Source
  • Only use the current IFU; older CDs carried an outdated version. Source

Sources

Where this note comes from.

Not covered: The PRO 6000 IFU was not opened, so display messages (such as position or cover errors), cleaning agents and battery life are missing; the one indicator listed comes from a recall notice. MAUDE has only two reports for the brand name.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.