Model note · Orthopedic Implants
Zimmer Persona The Personalized Knee System
Persona is Zimmer Biomet's main total knee system, built on its earlier NexGen line. It is a semi-constrained modular prosthesis that resurfaces femur, tibia and patella, with an anatomic, asymmetric tibial plate and many femoral sizes so the surgeon can match the patient closely. Options include Trabecular Metal porous components for cementless fixation, Ti-Nidium femurs for metal sensitive patients and Vivacit-E vitamin E polyethylene bearings.
The model
Maker, model and type.
- Maker
- Zimmer, Inc. (Zimmer Biomet)
- Model
- Persona The Personalized Knee System (total knee)
- Device type
- Orthopedic Implants
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Femur | Anatomic design, 21 profiles in 2 mm steps, standard and narrow, 12 A/P sizes, left and right | Source |
| Tibia | Anatomic, 9 sizes, left and right, stem compatible, Tivanium alloy | Source |
| Insert locking | Triple wedge locking mechanism, no lock down screws or through holes | Source |
| Bearings | 1 mm thickness steps, minimal to full constraint, conventional or Vitamin E polyethylene | Source |
| Ti-Nidium femur option | Titanium alloy (ASTM F136) femoral components with Ti-Nidium nitriding surface hardening | Source |
| Cementless option | Trabecular Metal femoral component allows fixation with or without cement | Source |
| Type | Semi-constrained modular knee prosthesis, modular between tibial plate and articular surface | Source |
| First US clearance | K113369, March 2012, product code JWH | Source |
Recalls
Known FDA recalls.
- Z-0410-20162015-11-24
EM proximal tube ratchet teeth not latching onto the rod.
FDA record - Z-0198-20152014-10-09
Persona stemmed 5 degree cemented tibia: cleaning process validation failure.
FDA record - Z-0123-20152014-10-07
Tibial articular surface inserter not heat treated, tip could deform.
FDA record - Z-0539-20132012-11-14
Cemented 5 degree stemmed natural tibia and related drill tools replaced to reduce tibial perforation risk.
FDA record
Sources
Where this note comes from.
- Zimmer Biomet Persona Knee product page
- FDA 510(k) K113369 summary, Zimmer Persona Knee System
- FDA 510(k) K122745 summary, Persona with Trabecular Metal femur
- FDA 510(k) K193223 summary, Persona Ti-Nidium CR femoral components
- openFDA recall search, Persona knee
- openFDA MAUDE problem counts, brand Persona
Not covered: An implant has no error codes. The Persona instructions for use and surgical technique were not opened, so MRI conditions, materials of each component and contraindications are incomplete. MAUDE counts under brand Persona mix total, partial and revision knees. Recalls for the separate Persona Partial Knee and Persona Revision lines, third party sensors (OrthoSensor VERASENSE) and stem extensions sold under other brands were excluded.