Skip to content

Model note · Orthopedic Implants

Zimmer Persona The Personalized Knee System

Persona is Zimmer Biomet's main total knee system, built on its earlier NexGen line. It is a semi-constrained modular prosthesis that resurfaces femur, tibia and patella, with an anatomic, asymmetric tibial plate and many femoral sizes so the surgeon can match the patient closely. Options include Trabecular Metal porous components for cementless fixation, Ti-Nidium femurs for metal sensitive patients and Vivacit-E vitamin E polyethylene bearings.

The model

Maker, model and type.

Maker
Zimmer, Inc. (Zimmer Biomet)
Model
Persona The Personalized Knee System (total knee)
Device type
Orthopedic Implants
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FemurAnatomic design, 21 profiles in 2 mm steps, standard and narrow, 12 A/P sizes, left and rightSource
TibiaAnatomic, 9 sizes, left and right, stem compatible, Tivanium alloySource
Insert lockingTriple wedge locking mechanism, no lock down screws or through holesSource
Bearings1 mm thickness steps, minimal to full constraint, conventional or Vitamin E polyethyleneSource
Ti-Nidium femur optionTitanium alloy (ASTM F136) femoral components with Ti-Nidium nitriding surface hardeningSource
Cementless optionTrabecular Metal femoral component allows fixation with or without cementSource
TypeSemi-constrained modular knee prosthesis, modular between tibial plate and articular surfaceSource
First US clearanceK113369, March 2012, product code JWHSource

Recalls

Known FDA recalls.

  • Z-0410-20162015-11-24

    EM proximal tube ratchet teeth not latching onto the rod.

    FDA record
  • Z-0198-20152014-10-09

    Persona stemmed 5 degree cemented tibia: cleaning process validation failure.

    FDA record
  • Z-0123-20152014-10-07

    Tibial articular surface inserter not heat treated, tip could deform.

    FDA record
  • Z-0539-20132012-11-14

    Cemented 5 degree stemmed natural tibia and related drill tools replaced to reduce tibial perforation risk.

    FDA record

Safety

Safety warnings from the record.

  • A tibial perforation acts as a stress riser and can lead to tibial fracture or tibial plate loosening. Source
  • Instruments that skipped hardening can deform during seating of the articular surface. Source

Sources

Where this note comes from.

Not covered: An implant has no error codes. The Persona instructions for use and surgical technique were not opened, so MRI conditions, materials of each component and contraindications are incomplete. MAUDE counts under brand Persona mix total, partial and revision knees. Recalls for the separate Persona Partial Knee and Persona Revision lines, third party sensors (OrthoSensor VERASENSE) and stem extensions sold under other brands were excluded.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.