Model note · Orthopedic Implants

Howmedica Osteonics Triathlon Total Knee System

Triathlon is Stryker's modular total knee replacement, with femoral, tibial and patellar components in cruciate retaining, cruciate substituting, posterior stabilised and total stabiliser versions. Its defining feature is a single radius femoral curve centred on the epicondylar axis, intended to keep the ligaments evenly tensioned through mid flexion. Stryker reports over 3 million implantations. For a technician, the active items are the reusable instrument trays and the traceability of implant lots rather than the implant itself.

The model

Maker, model and type.

Maker
Howmedica Osteonics Corp., dba Stryker Orthopaedics
Model
Triathlon Total Knee System (CR, CS, PS and TS options)
Device type
Orthopedic Implants
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Component materialsCobalt chromium alloy, titanium alloy, commercially pure titanium, UHMWPE and calcium phosphateSource
Femoral geometrySingle radius centred on the epicondylar axis, constant from 10 to 110 degrees of flexionSource
Femoral sizes8 sizes (1 to 8) in CR and PS, left and rightSource
X3 tibial insertsGUR1020 UHMWPE per ASTM F648, sequentially crosslinked and annealed, with a cobalt chromium locking wire per ASTM F90Source
Fixation optionsTritanium components cemented or uncemented; beaded with Peri-Apatite uncemented only; all polyethylene tibia cemented onlySource
Insert designsPS, CR, CS and TS tibial inserts; symmetric and asymmetric patellaeSource
MRIMR Conditional; scan parameters updated in the instructions for use under K242831Source
RegulatoryProduct codes JWH, MBH, KRR among others; latest labelling clearance K242831, December 2024Source

Recalls

Known FDA recalls.

  • Z-0164-20272026-09-16

    Reprocessing instructions gave a 30 minute dry time for TS baseplate finishing trays in blue wrap, which may leave residual moisture.

    FDA record
  • Z-0023-20252024-09-04

    Universal TS baseplate anterior locking tabs out of specification, possibly preventing insert locking wire engagement.

    FDA record
  • Z-0459-20222021-11-24

    Triathlon Pro tibial preparation tray kitted with the wrong cemented keel punch.

    FDA record
  • Z-0249-20162015-09-29

    MIS modular distal capture lots could disassociate during use.

    FDA record
  • Z-0399-20072006-12-04

    X3 CR tibial insert size 2 and size 7 packages swapped.

    FDA record
  • Z-0556-052004-10-27

    PS femoral component fatigue testing did not consistently meet the predicted load in the 510(k).

    FDA record

Safety

Safety warnings from the record.

  • Labelling now carries a contraindication for patients with sensitivity to the implant materials. Source
  • Use only the current MR Conditional parameters in the instructions for use when a patient with this knee is scanned. Source
  • Instrument mix ups in kitted trays can lead to unintended bone preparation if not caught before use. Source

Sources

Where this note comes from.

Not covered: An implant has no error codes. The full instructions for use, including exact MR Conditional parameters and the surgical technique, were not opened, so scan conditions are not stated. Size and geometry details come from a Stryker registry protocol rather than the current IFU. MAUDE terms mix clinical and device problems and are not failure rates. Over 40 Triathlon recall records exist, many for instruments; only a selection is listed.