Model note · Blood Pressure Monitor

Welch Allyn Connex Spot Monitor

The Connex Spot Monitor is a compact spot check vital signs monitor that takes noninvasive blood pressure, pulse rate, SpO2 and temperature for adult, paediatric and neonatal patients. It is used mainly on general wards, clinics and outpatient areas, often on a mobile stand or wall mount. Its distinguishing features are SureBP inflation step measurement, a choice of Nonin, Nellcor or Masimo SpO2, and direct sending of vitals to the electronic record over Wi-Fi, Ethernet or Bluetooth.

The model

Maker, model and type.

Maker
Welch Allyn, Inc. (Hillrom, now part of Baxter)
Model
Connex Spot Monitor
Device type
Blood Pressure Monitor
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ParametersNIBP, pulse rate, SpO2 and body temperature (normal and axillary modes); neonatal, paediatric and adultSource
Systolic and diastolic rangeSystolic 30 to 260 mmHg adult and paediatric, 20 to 120 mmHg neonate; diastolic 20 to 220 mmHg adult and paediatric, 10 to 110 mmHg neonateSource
Blood pressure determination timeTypically 15 seconds, maximum 150 secondsSource
Blood pressure accuracyMeets ANSI/AAMI SP10:2002 (mean error within 5 mmHg, standard deviation 8 mmHg)Source
Overpressure cutoffAdult and paediatric 300 mmHg plus or minus 15 mmHg; neonate 150 mmHg maximumSource
SureTemp Plus module26.7 to 43.3 C range, calibration accuracy plus or minus 0.1 C in Direct modeSource
Electrical100 to 240 V AC, 50 to 60 Hz; Class I, internally powered; type BF defibrillator proof applied parts; IPX1Source
Size and weight (standard chassis)16.1 x 23.4 x 5.8 cm, 1.3 kg with batterySource
DisplayColour touchscreen, 800 x 480 pixels, 530 cd/m2Source
Battery run time (2 cell pack)About 5.2 hours continuous; roughly 6.8 to 8.4 hours in the published spot check duty cyclesSource
ConnectivityWi-Fi, Ethernet, Bluetooth and Bluetooth Low Energy to the EHR; optional Masimo RRp respiration rateSource
FDA clearanceK142356, 21 CFR 870.2300, product code MWI, Class II, cleared 19 December 2014Source

Recalls

Known FDA recalls.

  • Z-2581-20232023-08-10

    Copper tape incorrectly placed on the back of the LCD; affected units were to be isolated and returned for replacement.

    FDA record
  • Z-1563-20252025-03-21

    Undeclared natural rubber band in the packaging of some reusable BP cuff configurations.

    FDA record

Safety

Safety warnings from the record.

  • Mishandled batteries can heat, smoke, explode or catch fire; do not short, crush, burn or open them. Source
  • Not for use near flammable anaesthetics, near MRI, in hyperbaric chambers, during transport outside the facility or on patients on heart lung machines. Source
  • With repeated Stat mode readings, check the limb regularly, as prolonged impaired circulation or a misplaced cuff can cause bruising. Source
  • Do not put the cuff on a limb with an IV infusion, compromised circulation, or the same side as a mastectomy, and avoid an SpO2 clip on the same limb. Source
  • Some reusable BP cuff packs shipped with the monitor contained an undeclared natural rubber band (recall Z-1563-2025), a risk for latex allergic patients. Source

Sources

Where this note comes from.

Not covered: The Connex Spot service manual is only mirrored on a site behind a bot check, so service level calibration steps, part numbers and the full disassembly procedure were not reviewed; error codes come from the 2017 user DFU only and the DFU revision should be matched to the unit's software. The 2017 DFU predates the Masimo RRp option.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.