GE HealthCare CARESCAPE V100 Vital Signs Monitor

The CARESCAPE V100 is a compact, battery powered vital signs monitor for adult, paediatric and neonatal spot checks and bedside monitoring outside intensive care. Its core is DINAMAP non invasive blood pressure (SuperSTAT, Classic or auscultatory calibrated), with optional Ohmeda, Nellcor or Masimo SpO2, Alaris or Exergen temperature and a built in printer. Alarms show as E codes in the systolic window.

The model

Maker, model and type.

Maker
GE Medical Systems Information Technologies, Inc. (GE HealthCare)
Model
CARESCAPE V100 Vital Signs Monitor
Device type
Blood Pressure Monitor
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Parameter optionsNIBP (SuperSTAT, Auscultatory or Classic); SpO2 (Ohmeda TruSignal, Nellcor OxiMax or Masimo SET); temperature (Alaris Turbo Temp, Tri-Site or Exergen TemporalScanner); printer optionalSource
NIBP accuracyMeets AAMI SP10 (mean error 5 mmHg, SD 8 mmHg); SP10:2002 for SuperSTAT, SP10:1992 for Classic and AuscultatorySource
Cuff pressure and overpressure cut offOperating 0 to 290 mmHg adult/paed, 0 to 145 mmHg neonate; cut off 300 to 330 mmHg adult/paed, 150 to 165 mmHg neonateSource
Maximum determination time120 s adult/paed; 85 s neonateSource
SuperSTAT BP range (adult/paed)Systolic 30 to 290, MAP 20 to 260, diastolic 10 to 220 mmHgSource
Battery6 V 3.3 Ah sealed lead acid; up to 11.5 h (NIBP every 15 min) or 5 h (heavy use); recharge about 5 h off, 8 h onSource
PowerExternal Class II converter, 100 to 240 VAC 0.5 A in, 12 VDC 1 A out; three internal fuses (DC input, battery, remote alarm)Source
Size and weight19.5 cm high; 21.9 cm wide (25.4 cm with Alaris temperature); 13.5 cm deep; 2.4 kg with batterySource
FDA clearanceK102426 (October 2010), predicate K073203; product codes MWI, DQA, DXN, FLL; Class II, 21 CFR 870.2300Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Do not test or service a sensor while it is monitoring a patient. Source
  • Only staff familiar with the monitor should use it, and it should be kept out of the patient's reach when idle. Source
  • The AC adapter has a non resettable, non replaceable fuse, so a failed adapter is replaced, not repaired. Source

Sources

Where this note comes from.

Not covered: The manual is a 2010 R1.5 edition hosted by a reseller; later revisions may differ. In the NIBP alarm table the code column and message column extracted separately, so the E89 to E80 pairing was read from table order and should be checked by a reviewer against a clean copy. The service manual (calibration, leak test, 900 series definitions) was not available. No FDA recall names the V100.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.