Model note · Blood Pressure Monitor
GE HealthCare CARESCAPE V100 Vital Signs Monitor
The CARESCAPE V100 is a compact, battery powered vital signs monitor for adult, paediatric and neonatal spot checks and bedside monitoring outside intensive care. Its core is DINAMAP non invasive blood pressure (SuperSTAT, Classic or auscultatory calibrated), with optional Ohmeda, Nellcor or Masimo SpO2, Alaris or Exergen temperature and a built in printer. Alarms show as E codes in the systolic window.
The model
Maker, model and type.
- Maker
- GE Medical Systems Information Technologies, Inc. (GE HealthCare)
- Model
- CARESCAPE V100 Vital Signs Monitor
- Device type
- Blood Pressure Monitor
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Parameter options | NIBP (SuperSTAT, Auscultatory or Classic); SpO2 (Ohmeda TruSignal, Nellcor OxiMax or Masimo SET); temperature (Alaris Turbo Temp, Tri-Site or Exergen TemporalScanner); printer optional | Source |
| NIBP accuracy | Meets AAMI SP10 (mean error 5 mmHg, SD 8 mmHg); SP10:2002 for SuperSTAT, SP10:1992 for Classic and Auscultatory | Source |
| Cuff pressure and overpressure cut off | Operating 0 to 290 mmHg adult/paed, 0 to 145 mmHg neonate; cut off 300 to 330 mmHg adult/paed, 150 to 165 mmHg neonate | Source |
| Maximum determination time | 120 s adult/paed; 85 s neonate | Source |
| SuperSTAT BP range (adult/paed) | Systolic 30 to 290, MAP 20 to 260, diastolic 10 to 220 mmHg | Source |
| Battery | 6 V 3.3 Ah sealed lead acid; up to 11.5 h (NIBP every 15 min) or 5 h (heavy use); recharge about 5 h off, 8 h on | Source |
| Power | External Class II converter, 100 to 240 VAC 0.5 A in, 12 VDC 1 A out; three internal fuses (DC input, battery, remote alarm) | Source |
| Size and weight | 19.5 cm high; 21.9 cm wide (25.4 cm with Alaris temperature); 13.5 cm deep; 2.4 kg with battery | Source |
| FDA clearance | K102426 (October 2010), predicate K073203; product codes MWI, DQA, DXN, FLL; Class II, 21 CFR 870.2300 | Source |
Recalls
Recalls.
This note lists no recalls. Check the FDA recall database for the latest record.
Safety
Safety warnings from the record.
- Do not test or service a sensor while it is monitoring a patient. Source
- Only staff familiar with the monitor should use it, and it should be kept out of the patient's reach when idle. Source
- The AC adapter has a non resettable, non replaceable fuse, so a failed adapter is replaced, not repaired. Source
Sources
Where this note comes from.
- GE CARESCAPE V100 Operator's Manual, software R1.5, 2048723-001A (April 2010), reseller copy on usme.com
- FDA 510(k) K102426 summary, CARESCAPE V100 Vital Signs Monitor
- openFDA MAUDE problem counts, CARESCAPE V100
Not covered: The manual is a 2010 R1.5 edition hosted by a reseller; later revisions may differ. In the NIBP alarm table the code column and message column extracted separately, so the E89 to E80 pairing was read from table order and should be checked by a reviewer against a clean copy. The service manual (calibration, leak test, 900 series definitions) was not available. No FDA recall names the V100.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.