Terumo Advanced Perfusion System 1

System 1 is a modular heart-lung machine used by perfusionists for cardiopulmonary bypass, arterial perfusion and regional perfusion for up to six hours. Roller pumps, centrifugal pumps, sensor modules and an electronic gas blender all sit on a CAN network and are set up and controlled from a 15 inch central touchscreen. Its defining traits are the distributed network design, configurable safety responses that link sensors to pumps, and an internal battery rated for at least 60 minutes.

The model

Maker, model and type.

Maker
Terumo Cardiovascular Systems Corporation
Model
Advanced Perfusion System 1 (System 1), bases 801763 and 801764
Device type
Cardiopulmonary Bypass
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Base voltage100/120 V or 220/240 V, 50/60 Hz (catalogue 801763 and 801764)Source
Base size and weight57.4 x 89.4 x 67.3 cm (H x W x D), 118.8 kgSource
Pumps supportedUp to eight pumps, including up to two centrifugal control unitsSource
Roller pumps4 inch head, 0 to 4 L/min; 6 inch head, 0 to 10 L/minSource
Centrifugal control unit range0 to 3,600 RPMSource
Monitoring modulesUp to 18 modules (pressure, temperature, flow, level, air bubble, occluder, interface)Source
Pressure module range-250 to 900 mmHg, two transducers per moduleSource
Air bubble detector sensitivity0.5 cc or larger up to 6 L/min on 3/8 inch tubing; 0.3 cc or larger up to 3 L/min on 1/4 inch tubingSource
Electronic gas blender0 to 10 L/min gas flow, FiO2 0.21 to 1.00Source
Battery backupAt least 60 minutes for a fully loaded system when fully charged; up to 13 hours to rechargeSource
Indicated durationUp to six hours of extracorporeal circulationSource

Recalls

Known FDA recalls.

  • Z-1108-20202020-01-10

    Roller pump shafts may fail at the locknut thread, preceded by atypical noise.

    FDA record
  • Z-0176-20192018-09-13

    EPGS service manual was not updated when software 1.30 was released.

    FDA record
  • Z-0435-20132012-11-14

    Reports of total loss of function where some units went blank.

    FDA record
  • Z-1320-20122012-03-16

    6 inch roller pump jammed during cardioplegia delivery in 27 reported instances.

    FDA record
  • Z-1224-20092009-01-12

    Central Control Monitor start-up failure, display and calibration problems, loss of function.

    FDA record
  • Z-0909-20092008-11-07

    Power manager board could short, causing loss of power, loss of monitor or arcing.

    FDA record

Safety

Safety warnings from the record.

  • A partly charged battery may give less than 60 minutes of backup for a fully loaded system. Source
  • Battery time falling faster than real time, or a red battery icon after returning to mains, indicates damaged or discharged batteries. Source
  • When pumps lose the central computer they keep running locally but may not respond to linked sensor alarms such as air or level. Source
  • Keep hand cranks and standby equipment ready, since several manual responses for pump errors rely on them. Source
  • The 6 inch roller pump has jammed during cardioplegia delivery in reported cases. Source

Sources

Where this note comes from.

Not covered: The service manual is not public, so internal fault codes and calibration procedures are not covered; error entries come from the public operator's manual. Many recalls are older than 2013 and may already be corrected on most installed units.