Skip to content

Model note · Cardiopulmonary Bypass

LivaNova Stockert Heater-Cooler System 3T

The Stockert 3T is a heater-cooler used with heart-lung machines. It circulates water through the oxygenator heat exchanger, the patient blanket and the cardioplegia heat exchanger to warm or cool the patient during bypass of 6 hours or less. It is not part of the blood path, but its water tank can grow nontuberculous mycobacteria (NTM) that can be carried by aerosol. Since 2017 LivaNova has issued deep cleaning services, validated disinfection instructions and a design upgrade. The 2020 Final Configuration (K191402) was tested to reduce transmission of aerosolised mycobacteria and draws tank emissions through a disposable aerosol collection set into the hospital vacuum.

The model

Maker, model and type.

Maker
LivaNova Deutschland GmbH (formerly Sorin Group Deutschland)
Model
Heater-Cooler System 3T (Final Configuration, with aerosol collection set)
Device type
Cardiopulmonary Bypass
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classification and clearancesCardiopulmonary bypass temperature controller, product code DWC; K052601 (2006), K191402 (25 February 2020, Final Configuration), K220635 (16 November 2022)Source
UseCirculates water through heat exchangers to warm or cool a patient during cardiopulmonary bypass lasting 6 hours or lessSource
Water circuitsThree circuits independent of the mains water supply: two patient circuits (blanket and oxygenator) fed from one tank, and one interchangeable cardioplegia circuit; circuits can be switched off separately to give more heating or cooling to the active onesSource
Aerosol collection setDisposable canister with a hydrophobic filter, connected between the 3T and the facility vacuum, drawing tank emissions under negative pressure; validated for a seven day use periodSource
Disinfection validationCleaning, disinfection and water preservation instructions validated to an intermediate level of disinfection and tested over 58 weeksSource
REF numbers in recalls16-02-85 (120 V, 60 Hz) and 16-02-81 (240 V)Source

Recalls

Known FDA recalls.

  • Z-1498-20202020-02-24

    Validated cleaning and disinfection instructions issued to reduce the risk of NTM contamination.

    FDA record
  • Z-0220-20202018-11-01

    Disinfectant and water preservative could react and become ineffective; design upgrade to reduce aerosol emission (8,671 units).

    FDA record
  • Z-1099-20192018-09-14

    Six units tested positive for NTM after deep cleaning.

    FDA record
  • Z-1901-20182018-04-20

    Deep cleaning service made available in the US for units suspected of contamination or visible biofilm.

    FDA record
  • Z-2771-20172017-06-30

    Hydrogen peroxide could diffuse across certain oxygenator heat exchanger fibres.

    FDA record

Safety

Safety warnings from the record.

  • NTM in the tank water can be spread as aerosol; the Final Configuration changes were made and tested to reduce transmission of aerosolised mycobacteria. Source
  • Hydrogen peroxide used in the water could diffuse through the fibres of certain oxygenators into the blood side above allowed limits (2017, 7,546 units). Source

Sources

Where this note comes from.

Not covered: The operating instructions were not read, so water volumes, temperature ranges, flow rates, the exact cleaning schedule and placement rules in theatre are not included. The FDA page lists references on Mycobacterium chimaera transmission from heater-cooler units; the clinical details (species, time to infection) should be added from those references before publishing.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.