Skip to content

Model note · ECG

Philips PageWriter TC70 cardiograph

The PageWriter TC70 is Philips' top cart based diagnostic resting ECG machine, used in emergency departments, cardiology and wards to record, interpret, print and send 12 to 18 lead ECGs. It pairs a 15 inch touchscreen with the Philips DXL algorithm, which adds STEMI aids such as ST maps and culprit artery statements, and it is built to push reports straight into ECG management systems over wired or wireless networks.

The model

Maker, model and type.

Maker
Philips Medizin Systeme Böblingen GmbH (Philips Healthcare)
Model
PageWriter TC70 cardiograph (product 860315)
Device type
ECG
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
LeadsUp to 18 simultaneously acquired leads; 20 minute full disclosure historySource
AcquisitionDefibrillation protected; 8,000 samples per second per lead wire; reports stored at 500 HzSource
Display15 inch TFT, 1024 x 768, five wire resistive touchscreenSource
PrinterThermal digital array, 200 dpi by 500 dpi at 25 mm/s; Z-fold letter or A4 paperSource
FiltersAC filter 50 or 60 Hz; high pass 0.05, 0.15, 0.5 Hz; low pass 40, 100, 150 HzSource
BatteryTwo hot swappable lithium ion modules (both required); about 11 hours or 80 printed reports per charge; under 8 hours to reach 90 percentSource
Power100 to 240 V AC, 50/60 HzSource
Storage200 ECGs internal, 200 more on optional USBSource
Size and weight330 x 405 x 135 mm; 13 kgSource
Operating environment10 to 40 C, 10 to 90 percent RH non condensing, up to 3,048 mSource
Connectivity10/100 Ethernet; 802.11 a/b/g/n with WPA2; PDF, Philips XML and DICOM outputSource

Recalls

Known FDA recalls.

  • Z-0853-20192019-01-09

    End of life batteries in PageWriter TC cardiographs may overheat, melt the case or ignite (TC70, software up to A.07.05.22).

    FDA record
  • Z-0857-20192019-01-09

    Same battery issue, TC70 government bundle with trolley.

    FDA record
  • Z-0858-20192019-01-09

    Same battery issue, TC70 government bundle without trolley.

    FDA record

Safety

Safety warnings from the record.

  • End of life batteries may overheat and set the cardiograph on fire, which can injure patients or staff nearby. Source
  • One MAUDE report describes a plugged in unit that sparked, caught fire and reignited after being extinguished. Source
  • Lead reversals can produce misleading interpretations, which is why the system checks for them before acquisition. Source

Sources

Where this note comes from.

Not covered: The PageWriter TC service manual and its battery addendum are not openly published by Philips, so no error or technical alert codes, part numbers or service tests are included; the 510(k) summary for the current release was not reviewed.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.