Philips Brilliance 64

The Philips Brilliance 64 is a whole-body 64-slice CT scanner with a continuously rotating tube and detector gantry, widely installed in hospitals for general, cardiac and emergency imaging. It is now an older platform that is often found on the refurbished market. Its long service life means collimator wear, couch mechanics and software level are frequent themes in its field history.

The model

Maker, model and type.

Maker
Philips Medical Systems (Cleveland), Inc.
Model
Brilliance 64 (model 728231)
Device type
CT Scanner
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Model number728231Source
Slices64Source
CoverageAbout 40 mm per pass, in about 0.4 sSource
X-ray tube familyPhilips MRC tubeSource
FDA clearance cited for the model510(k) K033326, product code JAK, regulation 21 CFR 892.1750, decision date 29 October 2003Source
Regulatory class (US)Class IISource

Recalls

Known FDA recalls.

  • Z-2191-20192019-06-24

    Cardiac step-and-shoot images could be labelled as phase tolerance applied when it was not.

    FDA record
  • Z-0543-20192018-10-18

    Fibreglass resin separation may let the gantry front cover and support fail or fall.

    FDA record
  • Z-1143-20192018-07-03

    Tube collimator compensator past expected life may fail, causing ring or smudge artefacts.

    FDA record
  • Z-1413-20182018-02-08

    Resetting the emergency stop may disable the horizontal motor so the couch top can move.

    FDA record
  • Z-2116-20162016-04-01

    Intermittent swirl-like ring artefacts on software version 3.5.4.

    FDA record
  • Z-1669-20112010-12-10

    Patient support vertical brake hub could disengage and allow vertical collapse.

    FDA record

Safety

Safety warnings from the record.

  • Until the software fix is fitted, resetting the emergency stop can leave the horizontal couch drive disabled so the couch top can move while a patient gets on or off. Source
  • A failed gantry front cover could fall, so unusual gantry noise is a reason to stop scanning. Source
  • With some software the DoseRight feature could suggest higher mAs than clinicians expect for large children, so paediatric protocols need checking. Source
  • A head holder on some configurations could contact the back wall of an ISO shelter while the couch moves through the bore. Source

Sources

Where this note comes from.

Not covered: Philips no longer hosts a Brilliance 64 product page or datasheet, so detector, generator and bore figures could not be confirmed from a manufacturer source; the slice and coverage figures come from a refurbisher review and should be checked. No public service manual or error code list was found. The 510(k) cited for the model in FDA records (K033326) describes the original Philips Plus platform.