Model note · Prosthetic Limbs

Össur RHEO KNEE XC microprocessor knee

The RHEO KNEE XC is a microprocessor controlled prosthetic knee for transfemoral and knee disarticulation users. Instead of hydraulics it uses a magnetorheological actuator: a fluid whose stiffness changes in a magnetic field, so the controller can vary resistance continuously through stance and swing. It is set up by the prosthetist with the Össur Logic app, and has a mechanical lock that can carry full weight if power is lost.

The model

Maker, model and type.

Maker
Össur hf.
Model
RHEO KNEE XC microprocessor knee
Device type
Prosthetic Limbs
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
ActuatorMagnetorheological actuator with mechanical lock (full extension, 10 or 20 degrees)Source
Weight limit136 kg; moderate to high impact, walking and occasional runningSource
Device weight1.6 kgSource
BatteryLithium ion, 2350 mAh, 33.84 Wh; about 3 hours to charge from empty; 48 to 72 hours per chargeSource
Power supplyFRIWO FW8030M/24, 100 to 240 VAC input, 24 VDC 1.25 A outputSource
Use temperature-10 to 40 C; charging 0 to 40 CSource
Environmental ratingWeatherproof: fresh water splashes allowed, no submersion, no salt or chlorinated waterSource
Service interval and lifeProfessional check recommended every 40 months; expected life 6 million steps, about 4 to 6 yearsSource
RF immunity10 V/m radiated (IEC 61000-4-3); keep portable RF equipment at least 30 cm awaySource

Recalls

Known FDA recalls.

  • Z-1789-20242024-03-25

    RHEO KNEE XC (RKNXC0003): firmware issues could cause unintended warnings and shutdown, with fall risk.

    FDA record
  • Z-1787-20242024-03-25

    RHEO KNEE (RKN130002): same firmware issue causing unintended warnings and shutdown.

    FDA record
  • Z-1322-20152015-02-26

    RHEO KNEE units released without completing the full assembly process, found in a service line audit.

    FDA record

Safety

Safety warnings from the record.

  • Unexpected power loss or battery depletion stops normal function and can cause a fall; users must recognise the power loss warnings. Source
  • Misreading of the user's activity by the knee increases the risk of falling. Source
  • Components from other makers are untested with this knee and risk structural failure. Source
  • Keep away from strong fields such as MRI and CT scanners, transformers and electrosurgical equipment. Source
  • Always unplug the power supply before donning; a damaged supply can cause electric shock. Source

Sources

Where this note comes from.

Not covered: The IFU used is an Össur document but was opened from a distributor's site; Össur's own download was not checked for a newer revision. No IP code is given, only a weatherproof description. The service manual, calibration procedure and any numbered fault codes for the prosthetist app are not public; error entries are the warning patterns in the user IFU. The manufacturerUrl product page was not opened. MAUDE numbers are very small and not failure rates.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.