Model note · IONM
Natus Xltek Protektor32 IOM System
The Xltek Protektor32 is an intraoperative neuromonitoring system that records EEG, evoked potentials and EMG and delivers transcranial motor evoked potential stimulation, run by Natus EPWorks software. It is built for spinal and neurosurgical cases, comes in laptop or cart configurations, and supports remote monitoring of several cases from one networked PC.
The model
Maker, model and type.
- Maker
- Natus Medical Incorporated (Excel-Tech Ltd, Xltek)
- Model
- Protektor32 IOM System (EPWorks software)
- Device type
- IONM
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Inputs and channels | 64 inputs, 32 channel multimodality monitoring | Source |
| Modalities | SEP, MEP, TcMEP, BAEP, VEP, EMG and EEG | Source |
| Electrical stimulator outputs | 16 high level and 2 low level programmable outputs | Source |
| Transcranial electrical MEP stimulation | Programmable, up to 1000 V and 1500 mA | Source |
| Other stimulators | Click and tone auditory stimulation, LED goggle flash stimulation | Source |
| Nerve integrity monitoring mode | Audible chimes for detected activity on up to 8 EMG channels | Source |
| External connections | Trigger in and trigger out for external devices | Source |
| FDA clearance | K093304, cleared 11 August 2010, product codes GWF, OLT, GWE, GWJ | Source |
Recalls
Known FDA recalls.
- Z-1067-20152015-01-02
EPWorks remote monitoring fault: remote user stopping free run waveforms gets an insufficient privilege message.
FDA record
Safety
Safety warnings from the record.
- The TcMEP mode is meant to detect acute corticospinal conduction problems from mechanical trauma or poor blood supply during surgery, and only staff trained in EEG, EP and EMG should interpret it. Source
- Transcranial stimulation reaches 1000 V and 1500 mA, so settings and electrode sites must follow the user manual. Source
Sources
Where this note comes from.
- Natus Xltek Protektor32 datasheet 006333C
- FDA 510(k) K093304 decision letter and indications, Protektor32
- FDA recall Z-1067-2015, EPWorks software in Protektor 32
- openFDA MAUDE problem counts, brand PROTEKTOR 32
Not covered: The datasheet is a 2014 Natus document hosted on a third party CDN, not natus.com; specs should be checked against the current manual. The user and service manual was only found on a reseller site that blocked access, so no error codes, part numbers or calibration steps are included. The public 510(k) PDF is a poor scan with no readable technical summary. MAUDE counts are very small.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.