Medtronic NIM Vital Nerve Integrity Monitor

The NIM Vital (console NIM4CM01, named NIM 4.0 in FDA recall records) is an intraoperative EMG nerve monitor used in thyroid, head and neck, ENT, spinal and peripheral surgery to locate and continuously check nerves at risk. It records EMG from muscles supplied by the nerve and alerts the surgeon visually and audibly when the nerve is stimulated or its response trends down. It adds a wireless patient interface, software electrosurgery muting and continuous Nervassure monitoring over its NIM 3.0 predicate.

The model

Maker, model and type.

Maker
Medtronic Xomed, Inc.
Model
NIM Vital Nerve Integrity Monitor (console NIM4CM01, listed by FDA as NIM 4.0)
Device type
IONM
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Main componentsTouchscreen console, patient interface (wired or wireless) and cart; uses NIM 3.0 disposables through adaptorsSource
Console power100 to 240 V plus or minus 10%, 50/60 Hz, with battery backupSource
Patient interface powerDC from the console when wired; internal battery when wirelessSource
Maximum stimulation current50 mA (NIM 3.0: 30 mA)Source
Electrosurgery mutingWired mute detector or software detection without a mute detectorSource
Monitoring modesIntermittent stimulation with NerveTrend reporting, Automatic Periodic Stimulation (APS) and continuous Nervassure monitoringSource
Compatible muting probeMetal Muting Probe, Ref 8220325, onlySource
FDA clearance510(k) K200759, decision 28 October 2020, 21 CFR 882.1870, product code GWF, Class II; predicate NIM 3.0 (K083124)Source
Maintenance intervalFunctional and safety checks at least once a yearSource

Recalls

Known FDA recalls.

  • Z-1623-20252025-03-10

    Certain software versions may produce increased stimulus artefact; fixed in software 1.7.5.

    FDA record
  • Z-2409-20242024-06-24

    Potential for false negative response; fixed in software 1.5.4.

    FDA record

Safety

Safety warnings from the record.

  • Paralysing anaesthetic agents that suppress EMG are contraindicated; muscle activity must return before monitoring is relied on. Source
  • Stray electrosurgical current can cause burns at other sites; avoid activating the ESU while the stimulator touches tissue or near electrodes. Source
  • Remove continuous monitoring electrodes before external defibrillation to prevent thermal injury. Source
  • Above 3 mA with Slim Prass or Prass Bipolar probes, current density may damage tissue. Source
  • A false negative can leave a nerve unprotected, so the surgeon must still rely on anatomy and other methods. Source

Sources

Where this note comes from.

Not covered: The NIM Vital user manual and service manual were not found publicly, so system error messages, channel count and the impedance limits are not listed. The 510(k) table gives stimulation figures in a garbled layout, so only the maximum current is quoted.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.