MicroTransponder Vivistim Paired VNS System

Vivistim is an implanted vagus nerve stimulator for chronic ischaemic stroke patients with moderate to severe arm impairment. Instead of running on a fixed duty cycle like epilepsy VNS, it delivers short bursts that a therapist triggers in time with rehabilitation movements, using a laptop application and a USB radio transmitter. Patients can also start home sessions by swiping a magnet over the generator.

The model

Maker, model and type.

Maker
MicroTransponder, Inc.
Model
Vivistim Paired VNS System (IPG 1001, Lead 3000, Wireless Transmitter 2000/2100, SAPS 4001)
Device type
VNS
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Output current0.0 to 3.5 mA in 0.1 mA steps; accuracy plus or minus 0.1 mA up to 1.0 mA, plus or minus 10% aboveSource
Frequency1, 2, 5, 10, 15, 20, 25 or 30 Hz, within 1%Source
Pulse width10 to 1000 microseconds, within 2 microseconds or 1%Source
Train duration0.5 to 60 sSource
Magnet activationMagnet field present for 10 ms plus or minus 1 ms starts stimulation 65 ms plus or minus 25 ms later; holding the magnet stops stimulationSource
TelemetryMICS band, about 403 MHz, up to 1 m; transmitter powered from laptop USB with a 2 m cableSource
LeadSilicone insulated, bifurcated bipolar lead with two helical nerve cuffs and an anchor tether; 2 mm and 3 mm cuffsSource
Battery voltage measurement accuracyPlus or minus 50 mVSource
FDA approvalPMA P210007, 27 August 2021, breakthrough device designationSource
Wireless transmitter updateModel 2000 replaced by Model 2100 for an obsolete part (supplement S001, 2022)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Contraindicated after bilateral or left cervical vagotomy. Source
  • Stimulation can cause hoarseness, coughing, swallowing difficulty or discomfort, and implantation can damage the vagus or nearby nerves. Source
  • The system is MRI Conditional, so scans need the labelled conditions to be checked first. Source

Sources

Where this note comes from.

Not covered: Battery capacity, expected longevity, IPG size and full MRI conditions are in the Vivistim labelling, which was not read. SAPS messages and error codes are not public. No recall matched Vivistim. The manufacturerUrl points to the FDA summary.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.

VNS