Model note · ECMO
Medtronic VitalFlow ECMO System console
VitalFlow is Medtronic's dedicated ECMO platform, cleared in the US for long term (over 6 hours) ECMO pumping in adults with acute respiratory or cardiopulmonary failure. A touchscreen console drives a centrifugal pump through a motor drive that FDA notes is identical to the drive used with Medtronic's Bio-Console 560. An ultrasonic probe measures flow and can detect bubbles, and an emergency hand crank can drive the pump if the console or motor fails.
The model
Maker, model and type.
- Maker
- Medtronic, Inc. (Medtronic Perfusion Systems; console originally from Michigan Critical Care Consultants, Inc., MC3)
- Model
- VitalFlow ECMO System console (REF 58100) with VitalFlow centrifugal pump and motor drive
- Device type
- ECMO
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| FDA classification | Blood pump for ECMO, long term (over 6 hours) use; 21 CFR 870.4100, Class II with special controls, product code QNR | Source |
| Clearances | K230364 VitalFlow Console (MC3), cleared 25 August 2023; K250199 upgrade over USB, decided 20 May 2025 | Source |
| Maximum flow | 7 L/min | Source |
| Maximum output pressure | 680 mmHg at 3800 RPM | Source |
| Flow and bubble sensing | Clamp on ultrasonic transit time flow probe, with optional bubble detection | Source |
| Drive | DC motor drive with microprocessor based controller; drive motor identical to the reference Medtronic External Drive Motor 560A used with Bio-Console 560 | Source |
| Backup drive | VitalFlow Emergency Handcrank to drive the pump manually if console or motor drive fails | Source |
| Data handling | Data download and streaming for circuit data only; no patient data stored; can display data from a compatible VitalFlow oxygenator | Source |
| Software update path | From K250199, field software upgrades over the USB port by Medtronic service only; older consoles needed each board (MC, SC, FPGA) updated with the case open | Source |
Recalls
Known FDA recalls.
- Z-0509-20262025-10-03
E70 error during normal use with the touch screen blank or unresponsive for up to two minutes.
FDA record
Safety
Safety warnings from the record.
Sources
Where this note comes from.
- Medtronic VitalFlow ECMO system brochure (US)
- FDA 510(k) K230364 summary, VitalFlow Console
- FDA 510(k) K250199 summary, VitalFlow Console (upgrade over USB)
- openFDA recall search, VitalFlow
- openFDA MAUDE problem counts, brand VitalFlow
Not covered: VitalFlow is newer (2023 in the US) than most ECMO consoles, so its field history is short. Medtronic's brochure blocked automated access, so no facts are taken from it; battery run time, weight, speed range and electrical ratings are therefore missing. The instructions for use with the full alarm and error list were not found publicly; only E70 from an FDA recall is listed. MAUDE counts mix console and disposable reports. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.