Abbott CentriMag Acute Circulatory Support System

CentriMag is Abbott's extracorporeal centrifugal blood pump system with a fully magnetically levitated rotor. It is used as the pump in ECMO circuits and for short term ventricular support. One console drives one disposable pump through a reusable motor, while an optional Mag Monitor gives a second display and can oversee two consoles in biventricular support. It is widely used in cardiac ICUs. Its recall history centres on the motor cable, pump locking, EMI and power surge immunity.

The model

Maker, model and type.

Maker
Abbott (formerly Thoratec Corporation; originally Levitronix LLC)
Model
CentriMag Acute Circulatory Support System (2nd Generation Primary Console, motor 102956, blood pump 102953)
Device type
ECMO
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
PumpSingle use polycarbonate centrifugal pump, fully magnetically levitated, no bearings or seals; 3/8 inch barbed inlet; priming volume 31 mLSource
Speed range0 to 5,500 RPM; typical operation 3,000 to 5,000 RPM; at least 1,000 RPM needed for forward flowSource
Flow and pressureUp to 10 L/min or a maximum pressure head of 600 mmHgSource
Flow probeClamp on ultrasonic em-tec adult flow probe, 3/8 x 3/32 inch tubing, optimised for 0 to 10 L/minSource
Pressure monitoringTwo pressure channels (P1, P2), range -150 to 900 mmHg, zeroable during supportSource
Console batteryInternal Li-ion pack; at least 120 minutes at 3,500 RPM and 5.5 L/min; recharge to 90% in 4 hours or less and 100% in 5 hours or less; at least 100 charge cyclesSource
PowerSingle phase AC mains for the 2nd Generation Primary ConsoleSource
Regulatory status510(k) for bypass under 6 hours and for ECMO circuits (product codes DWA, QNR; e.g. K234118 decided 26 January 2024); PMA P170038 for up to 30 days of post cardiotomy ventricular supportSource

Recalls

Known FDA recalls.

  • Z-1099-20262025-10-10

    Pump may not fit or lock securely into the motor, and the pump not inserted alarm may not trigger.

    FDA record
  • Z-2215-20242024-05-22

    2nd Generation console may not meet IEC 61000-4-5 and can shut down without alarm after a surge.

    FDA record
  • Z-0221-20202019-08-22

    Motor and pump problems caused by electromagnetic interference.

    FDA record
  • Z-0103-20192018-09-05

    Motor stops from wire breaks or shorts at the motor cable bend protection.

    FDA record
  • Z-1901-20082008-03-17

    Support could be interrupted when used with a Valleylab Force FX-C electrosurgical unit.

    FDA record

Safety

Safety warnings from the record.

  • A pump not locked into the motor may not alarm and can lead to delayed support, embolism, stroke or death. Source
  • After a large power surge an affected console may stop pumping completely with no alarm, so a backup system must be ready. Source
  • The system is contraindicated as a cardiotomy suction device and for patients who cannot take anticoagulation such as heparin. Source
  • The pump is afterload sensitive and preload dependent, so filling pressure and mean arterial pressure affect flow. Source

Sources

Where this note comes from.

Not covered: The CentriMag operating manual with its full alarm list is not publicly posted, so only one alarm named in an FDA recall is listed. Console dimensions, weight and electrical ratings were not found in an opened source. MAUDE counts include many reports with insufficient information and are not a failure rate. Recall text was read through openFDA because FDA recall pages returned errors at drafting time.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.