Mazor X Stealth Edition robotic guidance system

Mazor X Stealth Edition is Medtronic's spine surgery robot. It joins robotic trajectory guidance with Stealth optical navigation so the surgeon can place pedicle screws and instruments along a plan made on the patient's CT. A surgical arm mounted on the operating table holds a tool guide on the planned path, while a navigation camera tracks instruments. It is widely installed in spine centres, and its recalls are mostly software related.

The model

Maker, model and type.

Maker
Mazor Robotics Ltd. (Medtronic)
Model
Mazor X Stealth Edition robotic guidance system (REF TPL0059)
Device type
Surgical Robot
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FDA classificationStereotaxic instrument, 21 CFR 882.4560, Class II, product codes OLO (primary) and LLZSource
Latest clearanceK251316, Mazor X System / Mazor X Stealth Edition, decided 11 September 2025 (predicate K230064)Source
System componentsWorkstation with software, surgical system with arm and tool guide, navigation camera, accessories, instruments and disposable kitsSource
Navigation cameraNDI Vega Polaris optical camera, 75 kg, 1.9 x 0.65 x 0.61 mSource
Workstation1.85 x 1.20 x 0.64 m, 185 kgSource
Surgical system (arm unit)0.92 x 0.63 x 0.58 m, 25 kgSource
Stated accuracySystem mean accuracy under 1.5 mm; navigation mean positional error under 2 mm and mean trajectory error of 2 degreesSource
Power110 to 120 VAC 60 Hz, 220 to 240 VAC 50 Hz, or 100 VAC 50/60 Hz; maximum 1000 VA; IEC 60601-1 compliantSource
Imaging interfacesO-arm imaging system, 2D C-arm and 3D C-arm; converts 2D fluoroscopic projections into a 3D volumeSource
Integrated navigated drillsMedtronic Stealth-Midas and Stealth-Midas MR8 navigated drill systemsSource

Recalls

Known FDA recalls.

  • Z-1122-20262025-12-10

    Software errors that can cause incorrect surgical instrument positioning during spinal surgery.

    FDA record
  • Z-3156-20242024-07-17

    Lagging navigation display or mismatch between navigated screw and displayed screw information.

    FDA record
  • Z-0101-20242023-09-05

    Software correction for versions 4.2.2 and 5.0.1.

    FDA record
  • Z-2330-20202019-12-05

    Positioner Type II could let the surgical system detach from the operating table.

    FDA record

Safety

Safety warnings from the record.

  • Software errors in some versions can cause incorrect instrument positioning during spinal surgery; Medtronic issued mitigations to follow until the update is installed. Source
  • A surgical system that detaches from the table could fall onto the patient and cause blunt injury. Source
  • A mismatch between the navigated screw and the on screen screw information was reported in one software release. Source

Sources

Where this note comes from.

Not covered: No public Mazor X service manual or error code list was found, so errorCodes is empty. The Medtronic product page renders without readable content, so all specifications come from the FDA 510(k) summary. The Appendix A mitigations in the December 2025 recall notice were not reproduced in the FDA record. MAUDE counts are reports, not failure rates.