Medtronic ManoScan ESO high resolution manometry system

ManoScan ESO is a high resolution oesophageal manometry system used in GI physiology labs to measure the pressure and motor function of the oesophagus and its sphincters, for example in dysphagia, achalasia and pre-operative reflux work up. A solid state catheter with 36 circumferential sensors builds a continuous pressure topography map, which ManoView software analyses against the Chicago Classification.

The model

Maker, model and type.

Maker
Medtronic (Given Imaging Ltd., Yokneam, Israel)
Model
ManoScan ESO high resolution manometry system (ManoScan HRM modules with ESO catheters)
Device type
Oesophageal Manometry
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Standard catheter (MSC-1286)36 circumferential solid state channels, 4.2 mm diameter, 1.0 cm sensor spacingSource
Measurement points36 channels giving 432 points of measurementSource
Small diameter catheters2.75 mm with 7.5 mm (MSC-3886) or 10 mm (MSC-3887) spacingSource
Impedance catheter (MSC-3890-Z)36 pressure plus 18 impedance channels, 4.2 mmSource
HRM modulesA120 base module, A200 for impedance, A300 for 3D catheters, A400 for video inputSource
Module size and weightAbout 331 to 341 x 245.5 x 54.5 to 98.5 mm; 2.1 to 4.0 kg; safety class ISource
Power110 to 230 V AC, 50/60 HzSource
CalibrationMultiple point calibration for HRM catheters with monitoring of range and resolutionSource
SoftwareManoScan acquisition and ManoView ESO analysis; Chicago Classification v4.0 listed with ManoScan v3.1 and ManoView ESO v3.4Source
FDA clearance510(k) K151086 (ManoScan System), decision 28 April 2016, product code FFX, class II; predicate K091070Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Catheters have become stuck in the nose or oesophagus after studies, sometimes needing sedation or ENT and endoscopic removal and causing nasal trauma. Source
  • Several disinfectants, including Rapicide, Sterilox, hydrogen peroxide and Sporox, are not approved and may damage the catheter or change its performance. Source
  • Wrong timing of thermal compensation can move IRP across the 15 mmHg threshold and change the diagnosis in some studies. Source

Sources

Where this note comes from.

Not covered: No FDA recall for ManoScan was found, the ManoScan user guide (DOC-2551) and catheter IFU are not openly posted by Medtronic, so pressure range, accuracy, catheter life and any software error messages are not included.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.