Intuitive SureForm 60 / SureForm 45 Stapler and Reloads

SureForm is Intuitive's single use, fully wristed robotic stapler for the da Vinci X and Xi systems (IS4200 and IS4000). The surgeon clamps and fires it from the console, and each reload lays three staggered staple rows on each side of a knife cut. It comes in 60 mm and 45 mm lengths with straight and curved tip versions, plus an 8 mm SureForm 30 family, all limited to their own reloads.

The model

Maker, model and type.

Maker
Intuitive Surgical, Inc.
Model
SureForm 60 / SureForm 45 Stapler and Reloads
Device type
Surgical Stapler
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Compatible systemsda Vinci IS4000 (Xi) and IS4200 (X) onlySource
SureForm 60 reload staple line6 rows (3 each side of the cut), 90 staplesSource
SureForm 45 reload staple line6 rows, 66 staplesSource
Reload colours and unformed leg lengthWhite 2.5 mm, Blue 3.5 mm, Green 4.3 mm, Black 4.6 mmSource
Reload constructionSingle use cartridge with staples and a stainless steel knife, shipped sterile with a staple retainerSource
IndicationsResection, transection and anastomosis in general, thoracic, gynaecological, urological and paediatric surgery; buttress material allowedSource
FDA clearanceK173721 SureForm 60 (July 2018); K183224 SureForm 45 (January 2019); product codes NAY and GDW, 21 CFR 878.4750Source

Recalls

Known FDA recalls.

  • Z-1855-20262026-03-11

    8 mm SureForm 30 gray reloads for the curved tip stapler may give an incomplete staple line.

    FDA record
  • Z-0340-20232022-10-07

    SureForm 30 curved tip stapler and reloads: possible staple deployment failure and fragment generation.

    FDA record
  • Z-0999-20222022-03-17

    SureForm 45, 45 curved tip and 60: tissue pushout during firing; correction letter and user manual addendum.

    FDA record
  • Z-1890-20202019-07-01

    SureForm 60 green reloads: manufacturing variation may leave adjacent staples undeployed.

    FDA record
  • Z-1891-20202019-07-01

    SureForm 60 black reloads: same manufacturing variation.

    FDA record

Safety

Safety warnings from the record.

  • Tissue can be pushed out of the jaws during firing, mainly when crossing an existing staple line, leaving cut tissue that is not stapled. Source
  • Incomplete staple lines can cause air or anastomotic leaks that may need another procedure. Source
  • MAUDE also records detachment, material fragmentation and unintended system motion codes, so inspect the instrument after use and report fragments. Source

Sources

Where this note comes from.

Not covered: Intuitive's SureForm user manual and addendum were not openly available, so system messages, firing behaviour (such as automatic firing speed control), clamp limits and reprocessing details are not included and errorCodes is empty. Recall records were read through openFDA because FDA's recall pages were throttling automated requests. MAUDE counts for the brand are large and reflect both instrument and reload reports.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.