Model note · Surgical Stapler

Ethicon ECHELON FLEX Powered Plus articulating endoscopic linear cutter, 45 mm and 60 mm

The ECHELON FLEX family are single patient use articulating endoscopic linear cutters that lay six staggered rows of staples and cut between them, used in laparoscopic and open general, thoracic, gynaecological, urological and paediatric surgery, including lung, liver, pancreas and bowel work. The Powered Plus versions fire by a battery driven motor rather than hand force, and take ECHELON ENDOPATH reloads with Gripping Surface Technology in colour coded staple heights.

The model

Maker, model and type.

Maker
Ethicon Endo-Surgery / Ethicon US, LLC (Johnson & Johnson MedTech)
Model
ECHELON FLEX Powered Plus articulating endoscopic linear cutter, 45 mm and 60 mm (also manual ECHELON FLEX ENDOPATH models)
Device type
Surgical Stapler
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Staple and cut line45 mm model: about 45 mm staples, 42 mm cut; 60 mm model: about 60 mm staples, 57 mm cutSource
Staple rowsSix staggered rows, three each side of the cut lineSource
PowerPrimary lithium battery pack installed before use; motor driven firingSource
Shaft lengths (60 mm Powered Plus)280 mm (PCEE60A), 340 mm (PSEE60A), 440 mm (PLEE60A)Source
ReloadsECHELON ENDOPATH GST reloads in blue, gold, green, black and white (for example GST60B to GST60W)Source
Shaft movementShaft rotates freely both ways; distal articulation for lateral accessSource
SterilisationCobalt 60 irradiation, minimum 25 kGy, SAL 10^-6Source
FDA clearance510(k) K202665, decision 15 October 2020, 21 CFR 878.4750, product code GDW, class II; predicates K163454, K160521, K140560Source

Recalls

Known FDA recalls.

  • Z-1530-20132013-05-13

    ECHELON 60 mm black reload (ECR60T) could give incomplete staple lines from damage during firing.

    FDA record
  • Z-0200-20202019-10-03

    Out of specification anvil component in ECHELON FLEX ENDOPATH 60 mm (EC60A).

    FDA record
  • Z-0201-20202019-10-03

    Same anvil issue in Powered Plus 60 mm, 280 mm shaft (PCEE60A); related numbers Z-0202-2020 and Z-0203-2020 cover the 440 mm and 340 mm shafts.

    FDA record
  • Z-2326-20212021-06-03

    Powered Plus packs labelled 60 mm contained 45 mm devices.

    FDA record

Safety

Safety warnings from the record.

  • A compromised staple line can cause anastomotic leak, haemorrhage, further surgery or death. Source
  • Bunched tissue or obstructions such as clips or stents in the jaws can give an incomplete staple line. Source
  • Once the manual override has been used, the device is disabled for further firings and must be replaced. Source
  • Do not incinerate, autoclave, EtO sterilise or immerse the lithium battery pack, and keep it below 100 C. Source
  • Packs labelled 60 mm were found to contain 45 mm devices in one lot, so check device size at opening. Source

Sources

Where this note comes from.

Not covered: The stapler is single use with no service manual or error codes; the IFU reload chart with closed staple heights and the articulation angle were not retrieved, and MAUDE counts for the brand also include the powered vascular stapler and manual models.