GE HealthCare BiliSoft 2.0 Phototherapy System

BiliSoft 2.0 is a fibre optic blanket phototherapy system: an LED light box sends blue light through a flexible cable into a woven light pad placed under or around the baby. It is used in hospitals and, under professional instruction, at home, often alongside overhead lights. A magnetic pad connector, hour meter and front panel indicators for overheating and LED failure are the main serviceable features.

The model

Maker, model and type.

Maker
GE HealthCare (made with Lumitex, Inc.)
Model
BiliSoft 2.0 Phototherapy System
Device type
Phototherapy Unit
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Peak wavelength445 to 470 nmSource
Spectral irradiance as shipped (bare pad)Large pad 49 uW/cm2/nm plus or minus 25% (9 point check); small pad 70 uW/cm2/nm plus or minus 25% (6 point check)Source
Irradiance through pad cover or nestAbout 35 uW/cm2/nm (large) and 50 uW/cm2/nm (small)Source
Pad sizes and cableSmall 20 x 25 cm, large 25 x 30 cm light emitting area; cable 137 cm; pad under 0.5 kg; X-ray compatibleSource
Light box16.5 x 21 x 17.5 cm; under 1.7 kg; sound under 44 dB(A) at 1 mSource
Electrical100 to 240 V, 50/60 Hz, 0.25 to 0.75 A; Class II double insulated; type BF pad; touch current under 100 uA normal, 500 uA single faultSource
Ingress and environmentLight box IP21, pad IPX4; operating 5 to 35 C, 15 to 90% RH, 70 to 106 kPaSource
Service lifeLight box 7 years; light pad 3.5 years; LED module over 50,000 hours to a 25% dropSource

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • Infants need eye protection, and nearby babies' eyes should be shielded too; never look straight at the light. Source
  • Staff watching blue light for long periods may get headache, nausea or mild vertigo. Source
  • Monitor bilirubin during therapy; listed risks include apnoea, bronze baby syndrome, skin blistering and patent ductus arteriosus. Source
  • Use only manufacturer specified replacement parts. Source

Sources

Where this note comes from.

Not covered: The only 510(k) found is K053568 for the original Lumitex BiliSoft; no BiliSoft 2.0 specific summary was opened. The manual's troubleshooting table swaps the replacement thresholds (small under 27, large under 39) compared with its own note and output table (large 27, small 39); the note's values are used here and the discrepancy should be checked. No FDA recall names BiliSoft and MAUDE has only two reports.