Model note · Fetal Monitor

EDAN F9 / F9 Express Fetal & Maternal Monitor

The Edan F9 is a bedside cardiotocograph with a 12.1 inch screen for non-stress testing from 28 weeks and for labour monitoring. It measures fetal heart rate by pulsed Doppler or direct fetal ECG and uterine activity by toco or intrauterine catheter. The F9 Express adds maternal ECG, SpO2, NIBP and temperature, and both work with Edan's FTS-3 wireless telemetry transducers.

The model

Maker, model and type.

Maker
Edan Instruments, Inc.
Model
F9 / F9 Express Fetal & Maternal Monitor
Device type
Fetal Monitor
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
FHR techniquePulsed Doppler with autocorrelation processing, 2 kHz pulse repetition frequencySource
Ultrasound transducer12 crystals, effective radiating area 942 mm2 plus or minus 15% (F9 and F9 Express)Source
DECG and MECG heart rate range30 to 240 bpm; MECG rate averages the 12 latest RR intervalsSource
TOCO and IUPTOCO 0 to 100, resolution 1; IUP 0 to 100 mmHg, 5 uV/V/mmHg, auto zero or manualSource
Maternal SpO2 and NIBP (F9 Express)SpO2 50 to 100%, PR 30 to 240 bpm plus or minus 3 bpm; NIBP systolic 40 to 270 mmHg, diastolic 10 to 215 mmHg, mean error within plus or minus 5 mmHg, SD 8 mmHgSource
Recorder152 mm (GE) or 150 mm (Philips) paper; 1, 2 or 3 cm/minSource
Display, size and weight12.1 in LCD; 347 x 330 x 126 mm; about 5.5 kg (F9) or 6.3 kg (F9 Express)Source
Power100 to 240 VAC, 50/60 Hz, 1.0 to 0.5 A; rechargeable lithium ion batterySource
Ingress protectionMain unit IPX1 (recorder drawer shut, not wall mounted vertically); US and TOCO transducers IPX8Source
Applied part classificationFHR, TOCO, FM, IUP type BF; DECG type CF; SpO2 and NIBP BF defibrillator proof; ECG and TEMP CF defibrillator proofSource
EnvironmentOperating 5 to 40 C, 15 to 93% RH non condensing, 86 to 106 kPaSource
FDA clearanceK173042 (F6, F6 Express, F9, F9 Express; F2, F3), product code HGM, Class II, decision 30 August 2018Source

Recalls

Known FDA recalls.

  • Z-1141-20262025-10-24

    Cybersecurity labelling correction: a public IP address was used as the factory default and as a manual example; customers to check IP settings and file the manual corrigenda.

    FDA record

Safety

Safety warnings from the record.

  • The monitor is not suitable for use in the presence of flammable gases. Source
  • Only trained staff in antepartum, labour and delivery rooms should use it, and non-stress testing is cleared from 28 weeks of gestation. Source
  • A monitor left on a public default IP address is a network security exposure; quarantine affected stock until the correction is applied. Source
  • The main unit is only drip proof (IPX1); only the transducers are rated for immersion (IPX8). Source

Sources

Where this note comes from.

Not covered: Edan's own website did not return readable content and no public F9 user or service manual could be opened, so there are no error codes, alarm messages, part numbers or battery runtime; reseller battery figures conflict (2 h against 4 h) and were left out. Almost all facts come from one FDA 510(k) summary. MAUDE has only a handful of Edan perinatal reports, too few for a trend, and the battery entry is a generic wear item rather than evidence of failures.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.