Model note · EP Mapping System
Biosense Webster CARTO 3 EP Navigation System
CARTO 3 is a 3D electroanatomical mapping system used in electrophysiology labs to build real-time maps of the heart and show where catheters sit during arrhythmia diagnosis and ablation. It combines magnetic sensor location for the catheter tip with current-based Advanced Catheter Location (ACL) for electrodes along the shaft, which reduces reliance on fluoroscopy for catheter visualisation.
The model
Maker, model and type.
- Maker
- Biosense Webster, Inc. (Johnson & Johnson MedTech)
- Model
- CARTO 3 EP Navigation System
- Device type
- EP Mapping System
- Main document
- Open the public document
Key specifications
What the public record says.
| Item | Value | Source |
|---|---|---|
| Location technologies | Magnetic sensor technology (catheter tip) and Advanced Catheter Location, ACL (shaft electrodes) | Source |
| Hardware components | Patient Interface Unit (PIU), workstation with GUI, wide-screen monitors, intracardiac in and out ports, power supply, patches connection box and cables, pedals, Location Pad, Signal Processing Unit | Source |
| External reference patches | Six patches, three on the chest and three on the back | Source |
| Catalogue numbers | FG-5400-00, FG-5400-00U | Source |
| Regulatory class | FDA Class II programmable diagnostic computer, 21 CFR 870.1425, product code DQK | Source |
| Map types | Activation, propagation, potential and chamber geometry maps, plus body surface ECG and intracardiac electrograms | Source |
| Latest cleared version found | Software V9.0 with improved PIU Plus (K254085, January 2026) | Source |
Recalls
Known FDA recalls.
- Z-2369-20122009-11-17
Potential hazard when a Bloom EP stimulator was used with RF generators and other equipment on CARTO 3 V1.05.
FDA record - Z-0884-20112010-10-05
System interface cable could lose body surface ECG signals, causing inaccurate catheter locations on the 3D display.
FDA record - Z-1230-20122012-02-09
RF interference through some cable configurations caused jumping or disappearing catheter icons and error messages during ablation.
FDA record - Z-1045-20142014-02-12
CARTOUnivu module (V3.2.2 and V3.2.3) issue with Siemens Axiom Artis VB fluoroscopy systems in certain configurations.
FDA record - Z-0002-20162015-09-10
Ultrasound image disappeared when CARTO 3 needed restarting; affects the CARTOSOUND module.
FDA record - Z-2776-20162016-08-15
Correction to reinstall systems moved and installed in China without Biosense Webster testing and approval.
FDA record - Z-2880-20202020-06-22
VISITAG module software (V7.1.80) could lose tag sites during recalculation, risking extra ablation.
FDA record
Safety
Safety warnings from the record.
- A software defect in the VISITAG module (V7.1.80) could make ablation tags disappear during recalculation, leading to extra ablation and, rarely, perforation. Check the installed version against the correction. Source
- When the system needed restarting, the image disappeared from the connected ultrasound system in a reported case during pericardial effusion. Plan for imaging continuity if CARTO must be rebooted mid-case. Source
- A hazard was identified when a Bloom EP stimulator was used together with RF generators and other equipment on CARTO 3 V1.05; check approved equipment combinations. Source
- Catheter location data depend on specialised mapping catheters and reference devices; only use accessories specified for the system. Source
Sources
Where this note comes from.
- FDA 510(k) K072202 summary, original CARTO 3 System
- FDA 510(k) K221112 summary, CARTO 3 with Advanced Focus Mapping
- FDA 510(k) K254085 summary, CARTO 3 V9.0 with PIU Plus
- J&J MedTech press release, CARTO 3 Version 8
- openFDA MAUDE problem counts, brand CARTO 3
- openFDA recalls, product description CARTO 3
Not covered: No public operator or service manual was found, so error messages, location accuracy figures, power and environmental specs, and the service troubleshooting procedure are not included; no manufacturer product page for the base system was located.
BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.
EP Mapping System
Other models of this device.
Next