Biosense Webster CARTO 3 EP Navigation System

CARTO 3 is a 3D electroanatomical mapping system used in electrophysiology labs to build real-time maps of the heart and show where catheters sit during arrhythmia diagnosis and ablation. It combines magnetic sensor location for the catheter tip with current-based Advanced Catheter Location (ACL) for electrodes along the shaft, which reduces reliance on fluoroscopy for catheter visualisation.

The model

Maker, model and type.

Maker
Biosense Webster, Inc. (Johnson & Johnson MedTech)
Model
CARTO 3 EP Navigation System
Device type
EP Mapping System
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Location technologiesMagnetic sensor technology (catheter tip) and Advanced Catheter Location, ACL (shaft electrodes)Source
Hardware componentsPatient Interface Unit (PIU), workstation with GUI, wide-screen monitors, intracardiac in and out ports, power supply, patches connection box and cables, pedals, Location Pad, Signal Processing UnitSource
External reference patchesSix patches, three on the chest and three on the backSource
Catalogue numbersFG-5400-00, FG-5400-00USource
Regulatory classFDA Class II programmable diagnostic computer, 21 CFR 870.1425, product code DQKSource
Map typesActivation, propagation, potential and chamber geometry maps, plus body surface ECG and intracardiac electrogramsSource
Latest cleared version foundSoftware V9.0 with improved PIU Plus (K254085, January 2026)Source

Recalls

Known FDA recalls.

  • Z-2369-20122009-11-17

    Potential hazard when a Bloom EP stimulator was used with RF generators and other equipment on CARTO 3 V1.05.

    FDA record
  • Z-0884-20112010-10-05

    System interface cable could lose body surface ECG signals, causing inaccurate catheter locations on the 3D display.

    FDA record
  • Z-1230-20122012-02-09

    RF interference through some cable configurations caused jumping or disappearing catheter icons and error messages during ablation.

    FDA record
  • Z-1045-20142014-02-12

    CARTOUnivu module (V3.2.2 and V3.2.3) issue with Siemens Axiom Artis VB fluoroscopy systems in certain configurations.

    FDA record
  • Z-0002-20162015-09-10

    Ultrasound image disappeared when CARTO 3 needed restarting; affects the CARTOSOUND module.

    FDA record
  • Z-2776-20162016-08-15

    Correction to reinstall systems moved and installed in China without Biosense Webster testing and approval.

    FDA record
  • Z-2880-20202020-06-22

    VISITAG module software (V7.1.80) could lose tag sites during recalculation, risking extra ablation.

    FDA record

Safety

Safety warnings from the record.

  • A software defect in the VISITAG module (V7.1.80) could make ablation tags disappear during recalculation, leading to extra ablation and, rarely, perforation. Check the installed version against the correction. Source
  • When the system needed restarting, the image disappeared from the connected ultrasound system in a reported case during pericardial effusion. Plan for imaging continuity if CARTO must be rebooted mid-case. Source
  • A hazard was identified when a Bloom EP stimulator was used together with RF generators and other equipment on CARTO 3 V1.05; check approved equipment combinations. Source
  • Catheter location data depend on specialised mapping catheters and reference devices; only use accessories specified for the system. Source

Sources

Where this note comes from.

Not covered: No public operator or service manual was found, so error messages, location accuracy figures, power and environmental specs, and the service troubleshooting procedure are not included; no manufacturer product page for the base system was located.