Abbott EnSite X EP System

EnSite X is Abbott's 3D cardiac mapping and catheter navigation system, successor to EnSite Velocity and EnSite Precision. It shows the position of conventional and Sensor Enabled catheters inside the heart, builds a chamber model from collected points, and displays electrical activity as traces and as isopotential and isochronal maps. It can navigate in an impedance based mode or a magnetic primary mode, the latter needing catheters with magnetic sensors.

The model

Maker, model and type.

Maker
Abbott Medical (formerly St. Jude Medical)
Model
EnSite X EP System
Device type
EP Mapping System
Main document
Open the public document

Key specifications

What the public record says.

ItemValueSource
Navigation modesMagnetic primary mode or impedance based mode (the predicate had impedance only, with magnetic field scaling)Source
Map typesWaveform traces plus 3D isopotential and isochronal maps on a patient specific chamber surfaceSource
Patient connection modulesSurfaceLink Module, 20 pin and 80 pin Catheter Input Modules, and Direct Connect ports on the amplifierSource
Contact force optionContact Force Module shows force from compatible catheters when used with TactiSys Quartz equipmentSource
Compatible circular mapping catheter (example)Advisor VL SE: 4 F loop adjustable 15 to 25 mm, 10 or 20 electrodes, two impedance navigation electrodes and two magnetic sensorsSource
Software v1.1.1 changesUndriven patch input signals, composite respiration waveform, bio impedance scaling, respiration gating and metal distortion threshold updates; no hardware changeSource
FDA clearanceK202066 (2020) first clearance, product code DQK, 21 CFR 870.1425; reference device CARTO 3 (K180238)Source

Recalls

Recalls.

This note lists no recalls. Check the FDA recall database for the latest record.

Safety

Safety warnings from the record.

  • The model display must be used together with conventional EP techniques to confirm catheter position. Source
  • AutoMarks reflect user set stability and RF or PFA metrics only and do not show lesion effectiveness. Source
  • LivePoint displays should not be the only display of tissue proximity during a study. Source
  • Ignoring shift and drift detection messages can collect shifted data and contribute to model bloat, according to an Abbott MAUDE investigation. Source

Sources

Where this note comes from.

Not covered: Amplifier channel counts, field frame specifications, power requirements and the system's error messages are in the Instructions for Use on Abbott's manuals site, which was not read; no error codes are listed. No openFDA recall matched EnSite X. MAUDE counts mix event and malfunction reports and are not failure rates.

BMEssentia knowledge drafted from public information, reviewed by Hasan Khaled Ismail on 2026-10-24. Always follow the manufacturer's instructions for use and service manual.